3.
Start the Pre-use check by pressing Yes.
S
V
S
-5073_XX
4.
Follow the on-screen instructions.
5.
After pressing the touchpad Pre-use
check the message Do you want to start
a pre-use check? is displayed - confirm
by pressing Yes.
4.2.2 INTERNAL LEAKAGE TEST
SVS-0082_XX
Connect the test tube between the inspiratory
outlet and the expiratory inlet.
Important: Use only the MAQUET test tube.
4.2.3 BATTERY SWITCH TEST
The Pre-use check tests the ventilator’s ability
to switch between AC and battery power when
AC power is lost and restored:
When the on-screen instruction appears,
disconnect the ventilator from AC power.
When the on-screen instruction appears,
reconnect the ventilator to AC power.
Note: This test will not be performed if there
is less than 10 minutes of battery operation
available.
44
SERVO-s VENTILATOR SYSTEM V7.0, User's Manual
| 4 |
Operation overview
|
Summary of Contents for Maquet SERVO-s VENTILATOR SYSTEM V7.0
Page 1: ...x User s Manual SERVO s VENTILATOR SYSTEM V7 0 ...
Page 2: ......
Page 4: ...4 SERVO s VENTILATOR SYSTEM V7 0 User s Manual Infologic 1 64 2 1 TABLE OF CONTENTS ...
Page 14: ...14 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 1 Introduction ...
Page 40: ...40 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 3 Power supply ...
Page 66: ...66 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 5 Monitor and record ...
Page 118: ...118 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 6 Ventilation modes and functions ...
Page 162: ...162 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 11 Technical data ...
Page 176: ...176 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 13 Appendix User interface ...
Page 182: ...182 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 15 Index ...
Page 183: ...x ...
Page 185: ...SERVO s VENTILATOR SYSTEM V7 0 User s Manual ...