5.2.3 ADJUST SCALE/SWEEP SPEED
SVS-6008_XX
To set the sweep speed and amplitude for
displayed waveforms:
1.
Press the Quick access fixed key.
2.
Press the Waveform Scales touchpad.
3.
Press the touchpad corresponding to the
waveform whose scale you wish to
change or select a sweep speed (5, 10 or
20 mm/s).
4.
To adjust the scale of a waveform, turn
the Main Rotary Dial to the desired value
or use auto scale (press Auto).
Important: MAQUET does not recommend
using auto scale in Bi-Vent mode, when
patient breathing is spontaneous on both
levels.