6.4.3 VOLUME CONTROL IN DETAIL
1
2
3
4
SVS-9002_XX
1.
Volume Control assures a preset tidal
volume with constant flow during a preset
inspiratory time at a preset frequency.
2.
The inspiratory flow is constant and
depends on User Interface setting.
3.
Inspiration starts according to the preset
frequency or when the patient triggers.
4.
If the patient makes an inspiratory effort
during the inspiratory period, the ventilator
will switch to Pressure Support to satisfy
the patient´s flow demand.
Expiration starts:
When the preset tidal volume is delivered
and after the preset pause time.
When the flow returns to the set value after
the preset tidal volume is delivered and after
the preset pause time (on-demand support).
The patient is however always guaranteed
an expiration time corresponding to at least
20% of the total breath.
If the upper pressure limit is exceeded.
80
SERVO-s VENTILATOR SYSTEM V7.0, User's Manual
| 6 |
Ventilation, modes and functions
|
Summary of Contents for Maquet SERVO-s VENTILATOR SYSTEM V7.0
Page 1: ...x User s Manual SERVO s VENTILATOR SYSTEM V7 0 ...
Page 2: ......
Page 4: ...4 SERVO s VENTILATOR SYSTEM V7 0 User s Manual Infologic 1 64 2 1 TABLE OF CONTENTS ...
Page 14: ...14 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 1 Introduction ...
Page 40: ...40 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 3 Power supply ...
Page 66: ...66 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 5 Monitor and record ...
Page 118: ...118 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 6 Ventilation modes and functions ...
Page 162: ...162 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 11 Technical data ...
Page 176: ...176 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 13 Appendix User interface ...
Page 182: ...182 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 15 Index ...
Page 183: ...x ...
Page 185: ...SERVO s VENTILATOR SYSTEM V7 0 User s Manual ...