6.5.2 PRESSURE CONTROL IN DETAIL
1
2
3
SVS-9003_XX
1.
Pressure Control assures that the preset
inspiratory pressure level is kept constant
during the entire inspiration. Breaths are
delivered according to the preset
frequency, inspiration time and inspiratory
pressure level resulting in a decelerating
flow.
2.
The preset pressure level is controlled by
the ventilator. The resulting volume
depends on the set pressure level,
inspiration time and the patient´s lung
mechanical properties during each breath
with a decelerating flow.
3.
Inspiration starts according to the preset
frequency or when the patient triggers.
Expiration starts:
After the termination of preset inspiration
time
If the upper pressure limit is exceeded.
84
SERVO-s VENTILATOR SYSTEM V7.0, User's Manual
| 6 |
Ventilation, modes and functions
|
Summary of Contents for Maquet SERVO-s VENTILATOR SYSTEM V7.0
Page 1: ...x User s Manual SERVO s VENTILATOR SYSTEM V7 0 ...
Page 2: ......
Page 4: ...4 SERVO s VENTILATOR SYSTEM V7 0 User s Manual Infologic 1 64 2 1 TABLE OF CONTENTS ...
Page 14: ...14 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 1 Introduction ...
Page 40: ...40 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 3 Power supply ...
Page 66: ...66 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 5 Monitor and record ...
Page 118: ...118 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 6 Ventilation modes and functions ...
Page 162: ...162 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 11 Technical data ...
Page 176: ...176 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 13 Appendix User interface ...
Page 182: ...182 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 15 Index ...
Page 183: ...x ...
Page 185: ...SERVO s VENTILATOR SYSTEM V7 0 User s Manual ...