6.6 PRESSURE REGULATED
VOLUME CONTROL
6.6.1 FUNCTIONAL DESCRIPTION PRVC
The Pressure Regulated Volume Control
(PRVC) mode is a controlled breathing mode.
Breaths are delivered mandatorily to assure
preset volumes, with a constant inspiratory
pressure continuously adapting to the patient´s
condition. The flow pattern is decelerating.
SVS-5034_XX
SERVO-s Ventilator System can be configured
to set Tidal Volume or Minute Volume. The
following parameters are set:
1.
Tidal Volume (ml) or Minute Volume (l/min)
2.
Respiratory Rate (b/min)
3.
PEEP (cmH
2
O)
4.
Oxygen concentration (%)
5.
I:E ratio / Insp. time
6.
Inspiratory rise time (%/s)
7.
Trigg. Flow / Trigg. Pressure
SVS-5035_XX
The ventilator delivers a pre-set Tidal Volume.
The pressure is automatically regulated to
deliver the pre-set volume but limited to
5 cmH
2
O below the set upper pressure limit.
The flow during inspiration is decelerating. The
patient can trigger extra breaths.
86
SERVO-s VENTILATOR SYSTEM V7.0, User's Manual
| 6 |
Ventilation, modes and functions
|
Summary of Contents for Maquet SERVO-s VENTILATOR SYSTEM V7.0
Page 1: ...x User s Manual SERVO s VENTILATOR SYSTEM V7 0 ...
Page 2: ......
Page 4: ...4 SERVO s VENTILATOR SYSTEM V7 0 User s Manual Infologic 1 64 2 1 TABLE OF CONTENTS ...
Page 14: ...14 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 1 Introduction ...
Page 40: ...40 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 3 Power supply ...
Page 66: ...66 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 5 Monitor and record ...
Page 118: ...118 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 6 Ventilation modes and functions ...
Page 162: ...162 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 11 Technical data ...
Page 176: ...176 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 13 Appendix User interface ...
Page 182: ...182 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 15 Index ...
Page 183: ...x ...
Page 185: ...SERVO s VENTILATOR SYSTEM V7 0 User s Manual ...