6.6.2 PRVC IN DETAIL
1
2
3
SVS-9006_XX
1.
PRVC assures a set target minute
ventilation to the patient. The target
volume is based upon settings for Tidal
Volume, frequency and inspiration time.
2.
The inspiratory pressure level is constant
during each breath, but automatically
adapts in small increments
breath-by-breath to match the patient´s
lung mechanical properties for target
volume delivery.
3.
Inspiration starts according to a preset
frequency or when the patient triggers.
Expiration starts:
After the termination of preset inspiration
time
If the upper pressure limit is exceeded.
SERVO-s VENTILATOR SYSTEM V7.0, User's Manual
87
|
Ventilation, modes and functions
| 6 |