Always set the Apnea time appropriate to the
individual patient situation. If the apnea alarm
limit is reached the ventilator will automatically
switch to backup ventilation.
It is also very important to set lower and upper
alarm limit for expired Minute Volume.
It is important to set the Inspiratory rise time
to a comfortable value for the patient. In
Pressure Support the Inspiratory rise time
should normally be increased.
70%
Inspiratory Cycle-off is important for the
patient’s comfort and ventilator
synchronization with the patient. Inspiratory
Cycle-off is the point when inspiration
switches to expiration. E.g. for a patient with
expiratory resistance the inspiratory Cycle-off
should be set to a high value to guarantee
enough time for expiration.
Note: It is important to monitor the
corresponding Tidal Volume levels.
90
SERVO-s VENTILATOR SYSTEM V7.0, User's Manual
| 6 |
Ventilation, modes and functions
|
Summary of Contents for Maquet SERVO-s VENTILATOR SYSTEM V7.0
Page 1: ...x User s Manual SERVO s VENTILATOR SYSTEM V7 0 ...
Page 2: ......
Page 4: ...4 SERVO s VENTILATOR SYSTEM V7 0 User s Manual Infologic 1 64 2 1 TABLE OF CONTENTS ...
Page 14: ...14 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 1 Introduction ...
Page 40: ...40 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 3 Power supply ...
Page 66: ...66 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 5 Monitor and record ...
Page 118: ...118 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 6 Ventilation modes and functions ...
Page 162: ...162 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 11 Technical data ...
Page 176: ...176 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 13 Appendix User interface ...
Page 182: ...182 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 15 Index ...
Page 183: ...x ...
Page 185: ...SERVO s VENTILATOR SYSTEM V7 0 User s Manual ...