6.7.2 PRESSURE SUPPORT IN DETAIL
1
2
3
SVS-9004_XX
1.
Pressure Support assures that a preset
inspiratory pressure level is constantly
maintained upon patient effort.
2.
The preset pressure level is controlled by
the ventilator, while the patient determines
frequency and inspiration time.
3.
Inspiration starts when the patient triggers
a breath, gas flows into the lungs at a
constant pressure. Since the pressure
provided by the ventilator is constant, the
flow will decrease until the Inspiratory
Cycle-off is reached.
Expiration starts:
a. When the inspiratory flow decreases below
a preset fraction of the inspiratory peak
flow (Inspiratory cycle-off)
b. If the upper pressure limit is exceeded.
c. If the maximum time for inspiration is
exceeded.
d. If the flow drops to a flow range between
25% of the peak flow and lower limit for
Inspiratory Cycle-off fraction level and the
spent time within this range exceeds 50%
of the time spent in between the start of
the inspiration and entering this range.
SERVO-s VENTILATOR SYSTEM V7.0, User's Manual
91
|
Ventilation, modes and functions
| 6 |
Summary of Contents for Maquet SERVO-s VENTILATOR SYSTEM V7.0
Page 1: ...x User s Manual SERVO s VENTILATOR SYSTEM V7 0 ...
Page 2: ......
Page 4: ...4 SERVO s VENTILATOR SYSTEM V7 0 User s Manual Infologic 1 64 2 1 TABLE OF CONTENTS ...
Page 14: ...14 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 1 Introduction ...
Page 40: ...40 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 3 Power supply ...
Page 66: ...66 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 5 Monitor and record ...
Page 118: ...118 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 6 Ventilation modes and functions ...
Page 162: ...162 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 11 Technical data ...
Page 176: ...176 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 13 Appendix User interface ...
Page 182: ...182 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 15 Index ...
Page 183: ...x ...
Page 185: ...SERVO s VENTILATOR SYSTEM V7 0 User s Manual ...