6.10 BI-VENT
6.10.1 FUNCTIONAL DESCRIPTION
S
V
S
-5056_XX
Bi-Vent is pressure controlled breathing that
allows the patient the opportunity of
unrestricted spontaneous breathing. Two
pressure levels are set together with the
individually set duration of each level.
Spontaneous breathing efforts can be assisted
by pressure support
The following parameters are set:
1.
Pressure high (PHigh) for the higher
pressure level (cmH
2
O)
2.
PEEP for the lower pressure level (cmH
2
O)
3.
Oxygen concentration (%)
4.
Time at the higher pressure (THigh) level
(s)
5.
Time at the lower pressure (TPEEP) level
(s)
6.
Inspiratory rise time (s)
7.
Trigg. Flow / Trigg. Pressure
8.
Inspiratory Cycle-off (%)
9.
PS (Pressure Support level) above PHigh
(cmH
2
O)
10. PS (Pressure Support level) above PEEP
(cmH
2
O)
S
V
S
-5057_XX
SERVO-s VENTILATOR SYSTEM V7.0, User's Manual
103
|
Ventilation, modes and functions
| 6 |
Summary of Contents for Maquet SERVO-s VENTILATOR SYSTEM V7.0
Page 1: ...x User s Manual SERVO s VENTILATOR SYSTEM V7 0 ...
Page 2: ......
Page 4: ...4 SERVO s VENTILATOR SYSTEM V7 0 User s Manual Infologic 1 64 2 1 TABLE OF CONTENTS ...
Page 14: ...14 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 1 Introduction ...
Page 40: ...40 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 3 Power supply ...
Page 66: ...66 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 5 Monitor and record ...
Page 118: ...118 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 6 Ventilation modes and functions ...
Page 162: ...162 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 11 Technical data ...
Page 176: ...176 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 13 Appendix User interface ...
Page 182: ...182 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 15 Index ...
Page 183: ...x ...
Page 185: ...SERVO s VENTILATOR SYSTEM V7 0 User s Manual ...