7.3 ALARM SETTINGS FOR
BREATHING PARAMETERS
This section discusses viewing and setting
alarm limits, lists alarm settings, explains the
conditions under which alarm limits are set to
their default values.
See section Alarm limits on page 153.
7.3.1 VIEWING ALARM LIMITS
Alarm limits may be viewed in the Measured
Values Display on the right side of the screen.
See the Monitoring and Recording chapter for
details on the Measured Value Display.
7.3.2 SETTING ALARM LIMITS
To set alarm limits, touch the fixed key Alarm
Profile in the upper right corner of the screen
(see the Operation Overview chapter for details
about setting limits).
7.3.3 CONDITIONS LEADING TO DEFAULT
ALARM SETTINGS
Alarm limits become set to their default values
when:
restarting the ventilator
admitting a new patient
changing type of ventilation (option)
the ventilator has been totally without power
for more than 2 minutes.
7.3.4 LIST OF ALARM SETTINGS
Automatically Set
—These settings are
determined automatically by the ventilator
based on the related parameter settings:
O
2
concentration high (based on O
2
concentration setting)
O
2
concentration low (based on O
2
concentration setting)
High continuous pressure (based on PEEP
setting)
Upper Limit
—These settings define an upper
bound on a condition that is monitored by the
ventilator:
Paw high (airway pressure too high)
Apnea (maximum time exceeded)
Breathing Parameter Alarms
—These settings
define an allowed range for a breathing
parameter:
Expired minute volume (high and low)
Respiratory rate (high and low)
End Exp. Pressure (high and low)
126
SERVO-s VENTILATOR SYSTEM V7.0, User's Manual
| 7 |
Alarms
|
Summary of Contents for Maquet SERVO-s VENTILATOR SYSTEM V7.0
Page 1: ...x User s Manual SERVO s VENTILATOR SYSTEM V7 0 ...
Page 2: ......
Page 4: ...4 SERVO s VENTILATOR SYSTEM V7 0 User s Manual Infologic 1 64 2 1 TABLE OF CONTENTS ...
Page 14: ...14 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 1 Introduction ...
Page 40: ...40 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 3 Power supply ...
Page 66: ...66 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 5 Monitor and record ...
Page 118: ...118 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 6 Ventilation modes and functions ...
Page 162: ...162 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 11 Technical data ...
Page 176: ...176 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 13 Appendix User interface ...
Page 182: ...182 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 15 Index ...
Page 183: ...x ...
Page 185: ...SERVO s VENTILATOR SYSTEM V7 0 User s Manual ...