13.3 QUICK ACCESS KEY
This function allows the user to disconnect the patient from the ventilator and turns
off the alarms (for apnea, minute volume, frequency alarm, check tubing, low PEEP)
during suction support.
Note: Suction support is not available when NIV or O
2
breaths are activated.
Suction Support
Loops are graphs that show two measured values: one measured value (x-axis)
against another measured value (y-axis). Loops are updated breath by breath.
Two loops are available:
volume-pressure
flow-volume
Loops
Waveform scaling
Pressure scaling
Flow scale
Volume scale
These three scales are set to automatic scaling, by default.
The sweep speed (mm/s) can be set to 10 or 20 mm/s (default).
Waveform scales
Possibility to increase the space for viewing the waveform curves. This means that
more detailed information can be viewed.
Waveform
configuration
168
SERVO-s VENTILATOR SYSTEM V7.0, User's Manual
| 13 |
Appendix • User interface
|
Summary of Contents for Maquet SERVO-s VENTILATOR SYSTEM V7.0
Page 1: ...x User s Manual SERVO s VENTILATOR SYSTEM V7 0 ...
Page 2: ......
Page 4: ...4 SERVO s VENTILATOR SYSTEM V7 0 User s Manual Infologic 1 64 2 1 TABLE OF CONTENTS ...
Page 14: ...14 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 1 Introduction ...
Page 40: ...40 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 3 Power supply ...
Page 66: ...66 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 5 Monitor and record ...
Page 118: ...118 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 6 Ventilation modes and functions ...
Page 162: ...162 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 11 Technical data ...
Page 176: ...176 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 13 Appendix User interface ...
Page 182: ...182 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 15 Index ...
Page 183: ...x ...
Page 185: ...SERVO s VENTILATOR SYSTEM V7 0 User s Manual ...