11
Installation
Flusher disinfector S-406
9904
Checking and adjusting the door turning mechanism
May only be performed by authorized personnel.
This machine is connected to mains power supply and some
components are live.
If a program starts and the lid is obstructed by an object between the lid and the
top of the disinfector, the lid shall open again. The limit for this safety
precaution is between 6 and 12 mm as shown in figure 9.
Use the accompanying spacer bar to check this function.
• Stand the bar on edge (so that it is 12 mm high).
• Insert it between the top of the machine and the lid.
• Close the lid.
The lid must change direction and open again.
• Then turn the bar so that it is 6 mm high.
• Check to ensure that the lid closes.
Adjustment of the lid reversing function is described in the Service instructions.
V601
6-12 mm
Fig. 9. Closing the lid.