14
Flusher disinfector S-406
Installation
9904
Diameter of nozzle orifice at various steam pressures
If no steam pressure is specified, the disinfector nozzle is delivered with an
orifice diameter of 1.8 mm. If the pressure used is less than 300 kPa, the nozzle
orifice should be made larger as shown in the table below.
Steam pressure
kPa
60-80
80-125
125-200
200-250
250-300
5
4
3.5
3
2,5
Orifice diameter