background image

3

Installation

Flusher disinfector S-406

9904

Safety rules

This machine is designed with a number of integrated safety devices. To avoid
personal injury it is essential that the safety devices are not bypassed or in any
other way put out of action.

General safety rules

• The machine should be connected in accordance with the installation

instructions.

• The machine may not by used by juveniles.

• Installation and servicing should be carried out by staff with experience of

this machine.

• The lid lock of the machine may not be bypassed under any circumstances.

• Leakage in the system, eg, due to worn lid gasket, must be repaired

immediately.

• Personnel concerned must study valid handbooks and service manuals prior to

any repairs or service work.

• The machine may not be sprayed with water.

• Caution must be observed when using corrosive detergents.

• Safety precautions must be observed when using hot water or steam.

Power cut-off device

The machine must always be provided with a separate power cut-off device in
the power supply line, easily accessible on the wall.

Caution symbols

This manual contains certain warnings, instructions and advice of such
importance that they are particularly emphasized. The configuration and use of
the appertaining symbols are as follows:

This symbol indicates a warning in text included in the manual.
Such warnings refer to hazards that can result in more or less
serious injury and in some cases fatal injury.

It is also used in warnings to avoid damage to the machine.

Summary of Contents for S-406

Page 1: ...1 Installation Flusher disinfector S 406 9904 Flusher disinfector S 406 Serial No 203 238 4996959 01 INSTALLATION 0301 Produced 1998...

Page 2: ...4 Fixation in wall __________________________________________ 4 Connection of electricity water drain and possibly steam ________ 6 Goods arrangement chart __________________________________ 7 Deterge...

Page 3: ...ncerned must study valid handbooks and service manuals prior to any repairs or service work The machine may not be sprayed with water Caution must be observed when using corrosive detergents Safety pr...

Page 4: ...infector may be placed in a free standing position on four legs or mounted on a wall in a special wall bracket To facilitate cleaning a lateral distance of 100 mm to an adjacent wall or other equipmen...

Page 5: ...d remove the transport stay and attach the legs before fixing the disinfector in position Make sure that the disinfector hooks properly into the bracket when it is mounted on the wall fitting V300 Fig...

Page 6: ...l box can be unscrewed and pulled forward to facilitate installation Fig 4 Electrical connection see electric circuit diagram in Service instructions To prevent clogging of filters and valves flush th...

Page 7: ...art Place the notice with user s instruction on the wall above the disinfector Make sure that the lid does not obscure the chart Fig 5 Connection of water and steam supply Connect the disinfector to d...

Page 8: ...in the cans can be corrosive The container may also be placed on the floor or on the lower shelf of an adjacent sink unit up to a distance of about 70 cm from the disinfector The bottom of the contain...

Page 9: ...ld enter the service program as follows Simultaneously press 4 and 5 for 10 seconds The machine is then reset to the Service mode see also under the heading Service program The push buttons 1 5 will h...

Page 10: ...nt dosage systems two for detergent dosage and one for descaling compound For descaling and spray dosage the amount of compound can be set in the service program see Illustration 8 Set by the service...

Page 11: ...ector the lid shall open again The limit for this safety precaution is between 6 and 12 mm as shown in figure 9 Use the accompanying spacer bar to check this function Stand the bar on edge so that it...

Page 12: ...and apparatus with steam generator The final temperature is higher than 90 C The PROCESS COMPLETE lamp lights when the program is over Fig 10 Indicating lamps L M K Process in operation Process compl...

Page 13: ...0 C Noise level ca 59 dB A Power Model Electrical connection Fuse requirement L1 230 V 1N PE 50 Hz TN 1x16 A 2 94 kW L2 230 V 3 PE 50 Hz 3x16 A 3 91 kW L3 400 V 3N PE 50 Hz TN 3x10 A 3 91 kW L5 240 V...

Page 14: ...f no steam pressure is specified the disinfector nozzle is delivered with an orifice diameter of 1 8 mm If the pressure used is less than 300 kPa the nozzle orifice should be made larger as shown in t...

Page 15: ...15 Installation Flusher disinfector S 406 9904...

Page 16: ......

Page 17: ......

Page 18: ......

Page 19: ......

Page 20: ......

Page 21: ......

Page 22: ......

Page 23: ......

Page 24: ...This product is manufactured by GETINGE DISINFECTIONAB Ljungadalsgatan 11 Box 1505 351 15V xj Sweden...

Reviews: