9
Installation
Flusher disinfector S-406
9904
Settings and functional check
• Check to make sure that the disinfector is connected to the correct voltage as
indicated on the type data plate.
• Open the water valves, and steam valve if relevant, and turn on the main
switch.
To adapt the machine to the prevailing operating conditions, a number of
parameters must be checked and possibly reset when installing.
The following parameters are to be checked when installing:
• Amount of cleaning agent, see page 10.
• Amount of deliming agent, see page 10.
Checking and changing preset parameters
1
2
3
4
5
V310
To check the parameters for the above items, you should enter the service
program as follows:
• Simultaneously press
4
and
5
for 10 seconds.
The machine is then reset to the Service mode (see also under the heading
’Service program’). The push buttons
1
-
5
will have the following
functions:
1
Up
4
Abort
2
Down
5
Completed
3
ON/OFF
• Press
1
and
2
until you arrive at the desired line number as shown in
the table below:
Parameter
Line number
Basic setting
Setting range
Amount of cleaning agent *
42
25
0 - 99
Amount of deliming agent *
48
13
0 - 99
*
See further under the heading Setting the amount of detergent and descaling compound.
• Press 5 to enter the line. You can then read the set value and check it against
the table above.
N.B.
To change the contents of a line you must first key in your authority
code (1992) on line 00.
• Use
1
and
2
to change the contents of a line to the desired value as per
the above table.