background image

G E T IN G E S AF E S T EP AT P M O NI TO R

G E T IN G E S AF E S T EP AT P M O NI TO R

23

24

Symptom

Possible Causes

The display appears 

washed out or very 

dark

e Unit is too hot or too cold

e LCD contrast incorrectly adjusted 

     (see section 4.6)

e Unit is being used in inappropriate 

     lighting

f Unit or display damaged or faulty

Display segments 

missing or garbage 

displayed

e Display window is dirty

f Display window is scratched or dented

f Display or unit damaged or faulty

Keypad button has 

no effect when 

pressed

e Some buttons only work when electing 

     particular unit functions

e Lid not fully closed

f Keypad or unit damaged or faulty

Unit lid will not close 

properly or springs 

open during use

e Protective pocket incorrectly or not 

fully inserted

e RS232 connector cover is interfering 

     with the lid (see section 6.3)

e Test® Swab device incorrectly or 

     not fully inserted

f Lid or unit casework damaged

Measurement result 

is always zero RLU, or 

is much lower than 

expected

e Incorrect use of Test® Swab device

e Unit being used in an unstable

e Unit not in upright position when 

     measurement taken

e Test® Swab device is out-of-date

e Unit being used in an unstable thermal 

     environment – turn the unit off and 

     back on again

e Protective pocket or Test Swab device  

     not fully inserted into unit

e Protective pocket dirty or severely 

     scratched

 f Protective pocket damaged

 f Unit damaged or faulty

Symptom

Possible Causes

Measurement result 

is higher than ex-

pected

e Incorrect or inconsistent use of the 

Test® Swab device

e Lid not fully closed

e Unit being used in an unstable thermal 

    environment – turn the unit off and 

    back on again

f Unit casework or lid damaged or faulty

RS232 interface does 

not appear to be 

working

e Wrong interface cable being used

e Cable connector incorrectly inserted 

    into unit

e Cable connected to wrong port on PC

e Wrong PC software being used

e PC software incorrectly installed or 

    wrong options selected

f Cable or connectors damaged or 

    broken

f PC serial port or system software 

     faulty

f Unit damaged or faulty

7.3 Unit Error Codes

During normal operation, the unit performs various self-
checks on its internal components. If a problem is detected, 
the display will show an error number:

 

  

TIP: Most problems are likely to be transitory, and can be 
cleared by pressing the button, or by removing the batteries 
for 30 seconds and then reinserting them.

If any problem persists, please contact your Getinge 
representative for technical assistance, giving full details of 
the error code.

This section lists typical problems that might be 
encountered when using the unit, and their possible 
causes.

Some problems and causes can be rectified by the operator, 
while others may require technical assistant:

Severity

Action Required

e

This indicates a cause which can be rectified 

by the operator.

f

This indicates a cause which may require 

technical attention for rectification. Contact 

your Gteinge representative

7.1 Unit Beeps

During normal use, the unit emits a variety of different 
beeping sounds:

Beep Type

Possible Causes

Short high-pitched 

tone

e Unit turned on or off

e Sample measurement started

e Communications established with PC

e Results memory being erased

Long high-pitched 

tone

e Unit self-calibration complete

e Sample measurement complete

e Results memory erase complete

Long low-pitched 

tone

e Invalid date entered

e Program thresholds not set-up

e Invalid program threshold limits 

entered

e Attempting to select a non-failed

   (pass/caution) result for retesting

Two short high-

pitched tones

e Self-calibration required – remove

    Test® Swab device and close the lid

Three short high-

pitched tones

e Clock set-up required

e Batteries are flat

e Memory full warning

e Memory erase requested

e Unit error (see section 7.3)

7.2 Troubleshooting Tips

If the unit appears to be malfunctioning for any reason, 
carry out a thorough check for any obvious damage to the 
case, LCD display, lid, etc., caused by dropping or excessive 
physical mishandling.

The following table lists typical symptoms and their 
possible causes.

Symptom

Possible Causes

Unit will not turn on 

when the button is 

pressed

e Batteries are flat

e Batteries are the wrong type

e Batteries inserted incorrectly

f Unit or keypad damaged or faulty

Unit will not turn off 

when the button is 

pressed

e Unit is busy performing a reading or  

     self-calibration operation

e Unit lockup – remove the batteries for 

    10 seconds, then insert them again

f Unit or keypad damaged or faulty

Unit turns off unex-

pectedly

e Batteries are flat

e Batteries are loose within the battery 

     compartment

e Unit dropped or subjected to shock or 

     vibration

e Unit not used for 10 minutes and auto- 

     matically turns off into standby mode

f Unit damaged or faulty

Unit beeps when 

turned on, but noth-

ing is displayed

e Batteries are flat

f Unit or display damaged or faulty

The real time clock 

reverts to 12:00 

01/01/2000

e Batteries have been replaced

e Batteries are flat

e Batteries are loose within the battery 

     compartment

e Unit dropped or subjected to shock or 

     vibration

f Unit or display damaged or faulty

7. Troubleshooting

Summary of Contents for SafeStep

Page 1: ...Getinge SafeStep ATP Monitor Operator s Manual ...

Page 2: ...ling both during normal use and during storage and transportation CAUTION To avoid the risk of injury or choking never use the unit neck strap in situations where it could get caught in machinery etc Operating Precautions and Limitations of Use Batteries WARNING Use only non rechargeable alkaline batteries or rechargeable NiMH or NiCD batteries of types specified in section 10 WARNING Do not use b...

Page 3: ... EEC and 92 31 EEC as realigned by 93 69 EEC Regulatory Limitations of Use 1 INTRODUCTION 1 1 Key Unit Features 1 2 Accessories and Consumables 1 3 Principle of Operation 2 BASIC UNIT OPERATION 2 1 Unit Description 2 2 Keypad Symbols 2 3 Display Layout and Icons 2 4 Fitting the Batteries 2 5 Turning On the Unit 2 6 Low Battery Indicator 2 7 Internal Self Calibration 2 8 Turning the Unit Of 2 9 Pow...

Page 4: ...grammable users RS232 interface for results uploading to a PC Uses two standard or rechargeable batteries Replaceable protective sample pocket 1 Introduction 1 2 Accessories and Consumables Please contact your Getinge representative for details of unit accessories and consumables 1 3 Principle of Operation The Test Swab device uses a bioluminescent chemistry technology to convert an invisible conc...

Page 5: ...s and their definitions Low battery warning icon Lid icon flashes when the lid needs to be closed or the RS232 connector cover needs to be properly tucked under the lid Insert icon arrow flashes when the testing device should be inserted Remove icon arrow flashes when an testing device needs to be removed Pass result icon Caution result icon Fail result icon Sample measurement reading in Relative ...

Page 6: ... check sequence there must not be a Test Swab device in the unit and the lid must remain closed If the icon is shown with the arrow flashing open the lid and remove the Test Swab device from the unit If the icon is flashing close the lid TIP The Menu Select button can be pressed at any time to access the set up menu options Refer to section 4 for further details When the calibration self checks ar...

Page 7: ...on and Kit Inserts for full details of how to use the Test Swab and Test Sponge device NOTE Always allow the unit sufficient time to acclimatise to any change in environment temperature or humidity before taking sample measurements To perform a sample measurement follow the steps below For Flat Surfaces Using the Test Swab Swab the sample area and activate the Test Swab device see Kit Insert for c...

Page 8: ... button except the power button is pressed the countdown will restart To change the Standby Mode setting select the STANDBY MODE menu option by pressing the button Scroll to STANDBY MODE and press OK button Then use the up or down button to increase or decrease the Standby Mode duration followed by the OK button to store the new setting Standby Mode duration can be from 5 to 60 minutes in incremen...

Page 9: ...results a Program location can be selected for retesting by pressing the P button see section 4 2 With the unit turned on the Set up Menu can be accessed by pressing the button Use the up and down buttons to scroll through the menu list and then press the OK button to select the required menu option or the button the exit the menu The following 11 menu options are then available User ID LCD Contra...

Page 10: ...gured for either European format select D M Y for day month year or American format select M D Y for month day year D M Y or M D Y date D M Y or month M D Y month D M Y or date M D Y year TIP Pressing the button at any point will exit the clock set up mode leaving the time and date unchanged NOTE The clock does not have automatic daylight saving adjustment If this is required the time must be manu...

Page 11: ... unit 5 1 Conditions of Use As Getinge USA Inc has no control over the specification state or use of any computer equipment on which this software is installed or used the following Conditions of Use apply IMPORTANT The software is provided as is without warranties of any kind either expressed or implied including warranties of merchantability or fitness for a particular purpose Getinge USA Inc sh...

Page 12: ...he cable to your PC s USB port Step 9 Double click this icon to start the software Step 10 The first time you run the software the default user name Admin and password 1234 will be needed and may be changed in the software users form This is the default Admin account so the password should be changed if you choose so Step 11 Hit the Enter Key after you enter your login credentials If your ATP devi...

Page 13: ...reports and manage users and hand held devices You must be an admin to manage devices and users All users can perform upload and download functions You can Upload and Download data to and from the handheld to the database for reporting needs Click the upload button in the menu bar with the ATP device connected a With the handheld connected choose the Upload To Display button to see which program p...

Page 14: ...it using a wet cloth or by washing it under running water CAUTION Do not use solvents or other strong cleaning solutions as these may attack and deform the unit s plastic components and seriously degrade its performance 6 2 Replacing the Batteries For best results the batteries should replaced when the low battery warning icon is flashing Refer to section 2 4 for how to fit new batteries taking ca...

Page 15: ...technical assistance giving full details of the error code This section lists typical problems that might be encountered when using the unit and their possible causes Some problems and causes can be rectified by the operator while others may require technical assistant Severity Action Required e This indicates a cause which can be rectified by the operator f This indicates a cause which may requir...

Page 16: ...eader fault e Batteries are flat or loose f Unit s sample reader is damaged or faulty E9 Internal error e Batteries are flat or loose e Unit dropped or subjected to shock or vibration f Unit s damaged or faulty The Supplier warrants the Getinge SafeStep ATP Monitor unit when purchased new to be free from defects in materials and workmanship and will repair or replace at their discretion any SafeSt...

Page 17: ... Bus used to connect computer peripherals to a PC Unit The Getinge SafeStep ATP Monitor unit General Unit dimensions W x H x D 72mm x 191mm x 32mm Unit weight including batteries approx 260g Operating temperature range Relative Humidity range 5 C to 40 C 20 85 non condensing Storage temperature range Relative Humidity Range 10 C to 40 C 20 95 non condensing Unit Details Measurement range 0 to 9999...

Page 18: ...es in the United States or other countries Copyright 2017 Getinge AB or its subsidiaries or affiliates All rights reserved REVA Manufacturer Ruhof Corporation 393 Sagamore Ave Mineola NY 11501 1919 Sales Office Getinge 1777 East Henrietta Road Rochester NY 14623 www getinge com ...

Reviews: