background image

G E T IN G E S AF E S T EP AT P M O NI TO R

G E T IN G E S AF E S T EP AT P M O NI TO R

7

8

With the unit turned on, and having completed its internal 
calibration self-checks, it is then ready to perform a new 
sample measurement, and the display shows the following 
details:

Insert Sample Icon

Program Number    

 

 

Upper &   Lower  

and 

Location 

   Threshold 

Plan Number and Name 

 

 

User Name

Various keypad options are now available, as detailed in 
each of the following sections:

Button

Action

Section

Access set-up menu options

4

Select different User ID

4.1

Select program number

4.2

Select different Test Plan

4.3

Perform calibration

2.7

Start new measurement

3

View previous test results

4.5

Show statistical result data

4.6

Turn unit off

2.8

3.1 Programmable Test Locations

The unit can store up to 5000 individual programmable test 
locations (numbered PROG# 0 to PROG# 4999).

Each location is assigned a pair of upper ( A ) and lower ( B ) 
result threshold values, plus an optional location name.

When a sample measurement result is displayed, the 
RLU reading is compared against the selected Program 
thresholds to determine the overall Pass or Fail result:

Banding

Result

Reading ≤ lower threshold ( B )

e Pass

Reading > upper threshold ( A )

f Fail

For details of how to determine the appropriate program 
thresholds for your particular operating procedures, please 
contact your Getinge representative.

To set-up the Program thresholds, refer the section 4.2.

To select a Program location press the “P” button, then use 
the up and down buttons to change the Program PROG# to 
the desired Program location, followed by the “OK” button 
to accept the new location, or the “P” button to cancel the 
selection.

 

  

TIP: While selecting a PROG#, the display also shows the 
time and date that the Program location was last tested.

NOTE: If the selected Program location does not have 
any thresholds defined, they must be set-up using the 
PROGRAM menu option (see section 4.2) before the 
Program can be used.

3. Sample Measurements and 

Test Results

3.2 Taking a Sample Measurement

IMPORTANT: Please refer to the Test Swab and Test Sponge 
Product Specification and Kit Inserts for full details of how 
to use the Test Swab and Test Sponge device.

NOTE: Always allow the unit sufficient time to acclimatise 
to any change in environment, temperature or humidity 
before taking sample measurements.

To perform a sample measurement, follow the steps below:
For Flat Surfaces Using the Test® Swab
•  Swab the sample area and activate the Test® Swab device 

(see Kit Insert for complete useage details)

•  Open the unit lid, insert the Test® Swab device into the 

unit, and close the lid

•  Press the “OK” button and wait 15 seconds for the result 

to be displayed

For Scopes and Cannulated Instruments Using the Test 
Sponge and Test Swab
•  Pass Test Sponge through channel. (See Kit Insert for 

complete useage details)

•  Cut off tip of Test Sponge into the Test Swab tube
•  Swab outside surface of scope or instrument and activate 

Test® Swab device

•  Open the unit lid, insert the Test Swab device into the 

unit, and close the lid

•  Press the “OK” button and wait 15 seconds for the result 

to be displayed

WARNING: Always ensure that the exterior of the Test® 
Swab device is clean and dry before inserting it into the 
unit. Never insert anything other than a Test® Swab device 
into the unit. Never insert a device into the unit when the 
protective pocket is removed (refer to section 6.3).

While the measurement is being performed, the display 
counts down from 15 to 0:

 

  

NOTE: For consistent results, always keep the unit upright 
and steady while it is performing a measurement to ensure 
that the liquid in the Test® Swab device is at the bottom of 
the tube.

When the measurement is complete, the test result and 
overall pass/fail result are displayed:

Pass/Fail

Test number 

 

 

         Upper threshold

Program location   

 

         Lower threshold

Test time  

 

 

         Test date

TIP: A failed test result ( f ) can be selected for retesting 
by pressing the “P” button, followed by the “OK” button to 
select the Program location. See section 3.3 for details.

Now remove the Test® Swab device and dispose of it.

NOTE: For best results, and to prevent dust and dirt ingress, 
always keep the unit lid closed when not inserting or 
removing a Test® Swab device.

Previous test results can be viewed by pressing the up and 
down buttons (see section 4.5), or the results statistical 
data can be obtained by pressing the 

 button  

 (see section 4.6).

Summary of Contents for SafeStep

Page 1: ...Getinge SafeStep ATP Monitor Operator s Manual ...

Page 2: ...ling both during normal use and during storage and transportation CAUTION To avoid the risk of injury or choking never use the unit neck strap in situations where it could get caught in machinery etc Operating Precautions and Limitations of Use Batteries WARNING Use only non rechargeable alkaline batteries or rechargeable NiMH or NiCD batteries of types specified in section 10 WARNING Do not use b...

Page 3: ... EEC and 92 31 EEC as realigned by 93 69 EEC Regulatory Limitations of Use 1 INTRODUCTION 1 1 Key Unit Features 1 2 Accessories and Consumables 1 3 Principle of Operation 2 BASIC UNIT OPERATION 2 1 Unit Description 2 2 Keypad Symbols 2 3 Display Layout and Icons 2 4 Fitting the Batteries 2 5 Turning On the Unit 2 6 Low Battery Indicator 2 7 Internal Self Calibration 2 8 Turning the Unit Of 2 9 Pow...

Page 4: ...grammable users RS232 interface for results uploading to a PC Uses two standard or rechargeable batteries Replaceable protective sample pocket 1 Introduction 1 2 Accessories and Consumables Please contact your Getinge representative for details of unit accessories and consumables 1 3 Principle of Operation The Test Swab device uses a bioluminescent chemistry technology to convert an invisible conc...

Page 5: ...s and their definitions Low battery warning icon Lid icon flashes when the lid needs to be closed or the RS232 connector cover needs to be properly tucked under the lid Insert icon arrow flashes when the testing device should be inserted Remove icon arrow flashes when an testing device needs to be removed Pass result icon Caution result icon Fail result icon Sample measurement reading in Relative ...

Page 6: ... check sequence there must not be a Test Swab device in the unit and the lid must remain closed If the icon is shown with the arrow flashing open the lid and remove the Test Swab device from the unit If the icon is flashing close the lid TIP The Menu Select button can be pressed at any time to access the set up menu options Refer to section 4 for further details When the calibration self checks ar...

Page 7: ...on and Kit Inserts for full details of how to use the Test Swab and Test Sponge device NOTE Always allow the unit sufficient time to acclimatise to any change in environment temperature or humidity before taking sample measurements To perform a sample measurement follow the steps below For Flat Surfaces Using the Test Swab Swab the sample area and activate the Test Swab device see Kit Insert for c...

Page 8: ... button except the power button is pressed the countdown will restart To change the Standby Mode setting select the STANDBY MODE menu option by pressing the button Scroll to STANDBY MODE and press OK button Then use the up or down button to increase or decrease the Standby Mode duration followed by the OK button to store the new setting Standby Mode duration can be from 5 to 60 minutes in incremen...

Page 9: ...results a Program location can be selected for retesting by pressing the P button see section 4 2 With the unit turned on the Set up Menu can be accessed by pressing the button Use the up and down buttons to scroll through the menu list and then press the OK button to select the required menu option or the button the exit the menu The following 11 menu options are then available User ID LCD Contra...

Page 10: ...gured for either European format select D M Y for day month year or American format select M D Y for month day year D M Y or M D Y date D M Y or month M D Y month D M Y or date M D Y year TIP Pressing the button at any point will exit the clock set up mode leaving the time and date unchanged NOTE The clock does not have automatic daylight saving adjustment If this is required the time must be manu...

Page 11: ... unit 5 1 Conditions of Use As Getinge USA Inc has no control over the specification state or use of any computer equipment on which this software is installed or used the following Conditions of Use apply IMPORTANT The software is provided as is without warranties of any kind either expressed or implied including warranties of merchantability or fitness for a particular purpose Getinge USA Inc sh...

Page 12: ...he cable to your PC s USB port Step 9 Double click this icon to start the software Step 10 The first time you run the software the default user name Admin and password 1234 will be needed and may be changed in the software users form This is the default Admin account so the password should be changed if you choose so Step 11 Hit the Enter Key after you enter your login credentials If your ATP devi...

Page 13: ...reports and manage users and hand held devices You must be an admin to manage devices and users All users can perform upload and download functions You can Upload and Download data to and from the handheld to the database for reporting needs Click the upload button in the menu bar with the ATP device connected a With the handheld connected choose the Upload To Display button to see which program p...

Page 14: ...it using a wet cloth or by washing it under running water CAUTION Do not use solvents or other strong cleaning solutions as these may attack and deform the unit s plastic components and seriously degrade its performance 6 2 Replacing the Batteries For best results the batteries should replaced when the low battery warning icon is flashing Refer to section 2 4 for how to fit new batteries taking ca...

Page 15: ...technical assistance giving full details of the error code This section lists typical problems that might be encountered when using the unit and their possible causes Some problems and causes can be rectified by the operator while others may require technical assistant Severity Action Required e This indicates a cause which can be rectified by the operator f This indicates a cause which may requir...

Page 16: ...eader fault e Batteries are flat or loose f Unit s sample reader is damaged or faulty E9 Internal error e Batteries are flat or loose e Unit dropped or subjected to shock or vibration f Unit s damaged or faulty The Supplier warrants the Getinge SafeStep ATP Monitor unit when purchased new to be free from defects in materials and workmanship and will repair or replace at their discretion any SafeSt...

Page 17: ... Bus used to connect computer peripherals to a PC Unit The Getinge SafeStep ATP Monitor unit General Unit dimensions W x H x D 72mm x 191mm x 32mm Unit weight including batteries approx 260g Operating temperature range Relative Humidity range 5 C to 40 C 20 85 non condensing Storage temperature range Relative Humidity Range 10 C to 40 C 20 95 non condensing Unit Details Measurement range 0 to 9999...

Page 18: ...es in the United States or other countries Copyright 2017 Getinge AB or its subsidiaries or affiliates All rights reserved REVA Manufacturer Ruhof Corporation 393 Sagamore Ave Mineola NY 11501 1919 Sales Office Getinge 1777 East Henrietta Road Rochester NY 14623 www getinge com ...

Reviews: