GEX Doc# 100-166 Rev. B
Release Date: 05/24/2019
Page 2 of 3
Product Specifications and Usage Information
P4360
MS-2 Barcode Scanner
USAGE
1)
Installation
a.
The product includes the scanner with integrated USB cable and the screws and mount to secure it to the P4330.
When ordered with the P4330, it is typically attached and tested at GEX prior to shipment.
b.
Secure the scanner to the mount using the hex-head screws and 2.5mm driver tool (included). See Figure 1.
c.
Secure the mount assembly to the P4330 GEX Dosimeter Holder System baseplate using the provided hex-head
screws and 7/64” driver tool (included). See Figure 2.
FIGURE 1: MS-2 Barcode Scanner Assembly FIGURE 2: Secured Mount Assembly
d.
Insert the baseplate assembly into the sample compartment of the Evolution 220 using the registration pins at the
back, and then push down to secure the plate against the front slip-pin. See Figure 3.
e.
Ensure that the spring-loaded beam tubes fit flush against the walls. See Figure 4.
FIGURE 3: Registration pins and front slip-pin FIGURE 4: Spring-loaded beam tubes
f.
Carefully route the USB cord out the front of the sample compartment, through the flexible foam opening, and route
the cord to connect to the back of the PC. A USB extension cable may be required. See Figure 5 below.
2.
5m
m
7/
64”