Product Specifications and Usage Information 

P4400 

GENESYS 30 Spectrophotometer 

 

DESCRIPTION 

The P4400 Thermo Scientific GENESYS 30 Spectrophotometer (G30) is a device used to measure the optical absorbance of 

transparent samples in the spectrum of visible light.  It is designed to be simple and fast with the highest performance and 

quality.   
The spectrophotometer is built for the fast-paced, high-volume processing environment, and can withstand the rigors of daily 

usage for many years. 

APPLICATION 

The G30 is used for the measurement of optical absorbance of dosimeters using a select wavelength in the visible spectrum, 

such as GEX B3 radiochromic film dosimeters, and FWT-60 radiochromic film dosimeters.  The instrument is combined with 

specially designed dosimeter holders which position the dosimeters appropriately to obtain accurate and repeatable 

measurements.   

SPECIFICATIONS 

Physical Specifications: 

GEX Part 

No. 

Description 

Product Dimensions 

Packaging Dimensions 

Product 

Weight 

P4400-US 

Spectrophotometer – Type B plug 

34.3 cm L x 38.1 cm W x 20.3 

cm H  

(13.5” L x 15” W x 8” H) 

58.4 cm L x 50.8 cm W x 33.0 

cm H  

(23” L x 20” W x 13” H) 

10.0 kg  

(22 lbs.) 

P4400-EU 

Spectrophotometer – Type F Plug 

P4400-UK 

Spectrophotometer – Type G Plug 

 

**For a complete list of specifications see the brochures and information from Thermo Scientific. 

Material

 

Molded plastic

 

Color

 

Grey and White

 

Printing 

Thermo Scientific logo plate on front 

Electrical Supply 

100–240 V, 50–60 Hz, selected automatically; 150 W maximum 

Optical Design  

Single-beam  

Spectral Bandwidth(s)   

5 nm 

Light Source  

Tungsten-Halogen lamp 

Detector  

Silicon Photodiode 

Wavelength 

 

Range:    

325 –1100 nm 

 

Accuracy:  

±2 nm (full range 325 to 1100 nm) 

 

Repeatability: 

≤ 1

 nm  

Photometric 

Range:   

 

-3A to +3A 

 

Accuracy – Instrument: 

±0.002 A (0 – 0.3A) 

 

 

 

0.5% of ABS reading (0.301A – 2.5A) 

 

Repeatability: 

 

±0.002 A   

 

Drift (Stability) 

 

<0.002 A/hr 

 

Stray Light  

<0.1 %T at 340nm and 400nm 

Baseline Flatness 

<0.003 A 

Keypad 

Tactile rubber 23 keys with numeric keypad 

Display 

Touchscreen LCD panel; 800 x 480; 12.7 cm (5 in) diagonal 

 

 

Summary of Contents for P4400-EU

Page 1: ...Physical Specifications GEX Part No Description Product Dimensions Packaging Dimensions Product Weight P4400 US Spectrophotometer Type B plug 34 3 cm L x 38 1 cm W x 20 3 cm H 13 5 L x 15 W x 8 H 58 4...

Page 2: ...ge for dosimetry in radiation processing applications CAUTION Whenever the spectrophotometer has been stored or shipped immediate exposure to room air can cause condensation damage Move the shipping c...

Page 3: ...with the 2 5mm aperture film holder or any dosimeter sample in the sample compartment simply turn the unit off clear the obstruction and turn the instrument power back on 6 The instrument will perfor...

Page 4: ...ment is installed at the user s site The term Performance Verification replaces the older term Calibration Verification but the intent of either is the same to determine if the instrument is performin...

Page 5: ...range of wavelengths is the objective in particular for qualification and performance verification activities and this can be accomplished using the Scan feature of the GENESYS 30 Follow the procedure...

Page 6: ...col is subject to change without notice Users are ultimately responsible for validation of their systems GEX does not provide any warranty of any kind regarding this procedure IQ OQ PQ for use in B3 o...

Page 7: ...ntained and cared for will perform better and have a longer life Complete cleaning instructions which include the proper materials can be found in the GENESYS 30 User Guide Instructions for cleaning t...

Page 8: ...mm x 15mm GEX DoseStix style film dosimeters in the sample compartment for measurement P4420 WINdose Dosimeter holder Positioning 10mm x 10mm film dosimeters in the sample compartment for measurement...

Page 9: ...ar from the date of delivery for any customer dissatisfaction REFERENCES DOCUMENTS Thermo Scientific documents typically included with each instrument purchase GENESYS 30 User Guide GENESYS 30 Setup G...

Reviews: