background image

 

100-167 Rev. B 

Release Date: 10/29/18 

Page 

5

 of 

 

Product Specifications and Usage Information 

P4400 

GENESYS 30 Spectrophotometer 

 

 

FIGURE 4 

4.

 

Ensure the dosimeter holder is empty, then zero the instrument by pressing the yellow “0.00” button. 

5.

 

After the instrument is zeroed, insert a dosimeter to be measured and select the green diamond-shaped button.  The 

measurement will be captured on the screen.

  

See Figure 5 below.  Measurement time for each dosimeter is less than 

1 second.  

 

FIGURE 5 

6.

 

After the measurement is recorded, remove the dosimeter and insert the next sample to be measured.  Repeat step 5 

above for all dosimeters.   

NOTE:  Only 10 consecutive measurements may be taken.  If more than 10 measurements are needed, you must save or 

discard the data before proceeding.  The recorded measurements may be output to an external USB drive in the 

form of .CSV files. 

Scanning Measurements: 

Sometimes a series of measurements over a range of wavelengths is the objective, in particular for qualification and 

performance verification activities, and this can be accomplished using the Scan feature of the GENESYS 30. 
Follow the procedure below to determine the peak wavelength of an irradiated film dosimeter or reference standard 

(modify the procedure as necessary for other similar types of measurements): 
1.

 

On the home screen of the GENESYS 30 navigate to the scan icon. 

2.

 

Set the Scan Range (X-axis limits) at 500 – 600 nm in the absorbance mode by navigating using the arrow keys and 

typing in the values for the lower and upper limits. 

3.

 

Zero the GENESYS 30 with the intended dosimeter holder installed with no sample using the yellow 0.00 button. 

4.

 

When the zeroing process is complete insert a B3 dosimeter into the dosimeter holder. 

5.

 

Press the green Play (measure) button on the instrument keypad. 

6.

 

The resulting graph will display.  Using the arrow keys move the vertical line. 

7.

 

The resulting graph forms a peak, move the line until you reach the peak absorbance value.  Record the wavelength 

associated with this value; this is the peak absorbance wavelength. 

Summary of Contents for P4400-EU

Page 1: ...Physical Specifications GEX Part No Description Product Dimensions Packaging Dimensions Product Weight P4400 US Spectrophotometer Type B plug 34 3 cm L x 38 1 cm W x 20 3 cm H 13 5 L x 15 W x 8 H 58 4...

Page 2: ...ge for dosimetry in radiation processing applications CAUTION Whenever the spectrophotometer has been stored or shipped immediate exposure to room air can cause condensation damage Move the shipping c...

Page 3: ...with the 2 5mm aperture film holder or any dosimeter sample in the sample compartment simply turn the unit off clear the obstruction and turn the instrument power back on 6 The instrument will perfor...

Page 4: ...ment is installed at the user s site The term Performance Verification replaces the older term Calibration Verification but the intent of either is the same to determine if the instrument is performin...

Page 5: ...range of wavelengths is the objective in particular for qualification and performance verification activities and this can be accomplished using the Scan feature of the GENESYS 30 Follow the procedure...

Page 6: ...col is subject to change without notice Users are ultimately responsible for validation of their systems GEX does not provide any warranty of any kind regarding this procedure IQ OQ PQ for use in B3 o...

Page 7: ...ntained and cared for will perform better and have a longer life Complete cleaning instructions which include the proper materials can be found in the GENESYS 30 User Guide Instructions for cleaning t...

Page 8: ...mm x 15mm GEX DoseStix style film dosimeters in the sample compartment for measurement P4420 WINdose Dosimeter holder Positioning 10mm x 10mm film dosimeters in the sample compartment for measurement...

Page 9: ...ar from the date of delivery for any customer dissatisfaction REFERENCES DOCUMENTS Thermo Scientific documents typically included with each instrument purchase GENESYS 30 User Guide GENESYS 30 Setup G...

Reviews: