Product Specifications and Usage Information
P4400
GENESYS 30 Spectrophotometer
Notes:
•
GEX has completed the necessary testing on the instrument (described above) as part of the
DoseControl
System validation. The validation data is available for audit.
•
It is suggested to conduct thirty (30) repeated measurements of single dosimeters irradiated at low,
middle, and high doses in the user’s range, and to reproduce this test for a minimum of three (3) days
using 3 or more operators.
•
Some may wish to perform a formal Measurement Systems Analysis (MSA) and redefine the steps
suggested herein.
CARE & MAINTENANCE
Cleaning:
The instrument’s exterior and the sample compartment should be periodically cleaned as part of a Preventative Maintenance
program along with the dosimeter holder system. An instrument that is maintained and cared for will perform better and have
a longer life. Complete cleaning instructions which include the proper materials can be found in the
GENESYS 30
User Guide.
Instructions for cleaning the dosimeter holder system may be found separately in
GEX Doc# 100-168, GENESYS 30 Dosimeter
Holder System - Product Specifications and Usage (PSU)
.
Manufacturer Service:
If the device requires service for any reason there are two options:
1.
Depot Service:
For customers in the USA and Canada, the GENESYS 30 needs to be sent to Thermo Scientific in Madison, WI, U.S.A.
for repair or other services. Please contact GEX Customer Service at
for more information if service
is required.
2.
On-site Service:
On-site service may be available in some countries. Contact GEX Customer Service at
or your local
Thermo Scientific representative for details.
Lamp Replacement:
See GEX Doc# 100-169, GENESYS 30 Replacement Lamp - Product Specifications and Usage (PSU)
for complete details about the
lamp. The procedure for replacement of the lamp is below: