background image

JB0112-016

User Manual

Federal law (USA) restricts this device to sale by or on the order of a physician. This
product should not be used unless the operator has been instructed by a qualified
healthcare professional.

This device is for use by one person only; in order to avoid infection, do not share it.

NOTE: Check ALL parts for shipping damage. If shipping damage is noted, DO NOT use.
Contact your Graham-Field authorized distributor for further instruction.

NOTE: Do not use the Aspirator without first reading and understanding this manual! Save this
manual for future use.

JB0112-016-INS-LAB-RevB10

Medical Equipment

Complies with EN60601-1 standards

Summary of Contents for JB0112-016

Page 1: ...ice is for use by one person only in order to avoid infection do not share it NOTE Check ALL parts for shipping damage If shipping damage is noted DO NOT use Contact your Graham Field authorized distr...

Page 2: ...ts reserved When using electrical products especially when children are present basic safety precautions should always be followed Read all instructions before using Please note the following special...

Page 3: ...THE RISK OF ELECTRICAL SHOCK OR PERSONAL INJURY DO NOT DISASSEMBLE THIS UNIT Disassembly or attempted repairs if accomplished incorrectly can create electrical shock hazard Only qualified technicians...

Page 4: ...and or tissue damage This product should not be used unless the operator has been instructed by a physician or another qualified healthcare professional 1 Familiarize yourself with this unit 2 Place...

Page 5: ...specification label on the side of the aspirator to ensure that the voltage and current indicated on the unit match the voltage and current available Make sure the ON OFF switch shown in Figure 2 is...

Page 6: ...rease vacuum and counter clockwise to decrease vacuum Refer to the vacuum gauge while setting the desired level of vacuum To accurately read the gauge block the patient end of the hose or cap off the...

Page 7: ...been damaged 8 CLEANING INSTRUCTIONS Clean Collection Bottle 1 Turn off the aspirator allow vacuum to drop then disconnect AC power source 2 With the collection bottle still in the holder remove the...

Page 8: ...Replace with a new John Bunn bacteria HEPA filter and connect it to the tubing connected to the vacuum gauge and the tubing connected to the collection bottle lid assembly as formerly positioned Addi...

Page 9: ...or these possible causes PROBLEM ACTION Unit does not turn on 1 Check power source and connection 2 Ensure that the wall outlet is live 3 Check power cord for damage Motor runs but no vacuum 1 Verify...

Page 10: ...ons Collection Bottle Capacity 800 cc Operating Temperature 32 104 F 0 40 C Operating Humidity 95 RH Operating Atmospheric Pressure 10 2 15 4 psi 70 106 kP a Storage Temperature 14 158 F 10 70 C Stora...

Page 11: ...ontact GF Health Products Inc at the address below DO NOT return products to our factory without prior authorization LIMITATIONS AND EXCLUSIONS The foregoing warranty excludes consumables such as filt...