background image

Contents

1. Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
2. AC Power Source  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
3. Electrical  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
4. Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
5. Illustration of Important Parts  . . . . . . . . . . . . . . . . . . . . . . . . . . .5
6. Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
7. How to Operate Your John Bunn

®

Aspirator  . . . . . . . . . . . . . . .6

8. Cleaning Instructions  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
9. Maintenance  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8

10.  Troubleshooting  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
11.  Specifications / Classification . . . . . . . . . . . . . . . . . . . . . . . . . .10
12.  Limited Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11

2

JB0112-016-INS-LAB-RevB10

Graham-Field

and 

John Bunn

are registered trademarks, and 

Vacutec

is a trademark, of GF Health Pr oducts, Inc.

Packaging, warranties, products, and specifications are subject to change without notice. GF Health Pr oducts, Inc. is
not responsible for typographical errors.

© 2010 GF Health Products, Inc. All rights reserved.

When using electrical products, especially when children are present, basic safety precautions
should always be followed. Read all instructions before using. Please note the following special
statements, used throughout this manual, and their significance:

DANGER:

Indicates an imminent hazard situation that, if not avoided, will result in death or

serious personal injury.

WARNING:

Indicates a potential hazard situation that, if not avoided, could result in death

or serious personal injury.

CAUTION:

Indicates a potential hazard or unsafe practice that, if not avoided, could result

in product or property damage.

NOTE: Provides application recommendations or other useful information to ensure that you get
the most from your product.

READ THIS ENTIRE MANUAL BEFORE USING THE ASPIRATOR.

SAVE THIS MANUAL

1. SAFETY INFORMATION

Summary of Contents for JB0112-016

Page 1: ...ice is for use by one person only in order to avoid infection do not share it NOTE Check ALL parts for shipping damage If shipping damage is noted DO NOT use Contact your Graham Field authorized distr...

Page 2: ...ts reserved When using electrical products especially when children are present basic safety precautions should always be followed Read all instructions before using Please note the following special...

Page 3: ...THE RISK OF ELECTRICAL SHOCK OR PERSONAL INJURY DO NOT DISASSEMBLE THIS UNIT Disassembly or attempted repairs if accomplished incorrectly can create electrical shock hazard Only qualified technicians...

Page 4: ...and or tissue damage This product should not be used unless the operator has been instructed by a physician or another qualified healthcare professional 1 Familiarize yourself with this unit 2 Place...

Page 5: ...specification label on the side of the aspirator to ensure that the voltage and current indicated on the unit match the voltage and current available Make sure the ON OFF switch shown in Figure 2 is...

Page 6: ...rease vacuum and counter clockwise to decrease vacuum Refer to the vacuum gauge while setting the desired level of vacuum To accurately read the gauge block the patient end of the hose or cap off the...

Page 7: ...been damaged 8 CLEANING INSTRUCTIONS Clean Collection Bottle 1 Turn off the aspirator allow vacuum to drop then disconnect AC power source 2 With the collection bottle still in the holder remove the...

Page 8: ...Replace with a new John Bunn bacteria HEPA filter and connect it to the tubing connected to the vacuum gauge and the tubing connected to the collection bottle lid assembly as formerly positioned Addi...

Page 9: ...or these possible causes PROBLEM ACTION Unit does not turn on 1 Check power source and connection 2 Ensure that the wall outlet is live 3 Check power cord for damage Motor runs but no vacuum 1 Verify...

Page 10: ...ons Collection Bottle Capacity 800 cc Operating Temperature 32 104 F 0 40 C Operating Humidity 95 RH Operating Atmospheric Pressure 10 2 15 4 psi 70 106 kP a Storage Temperature 14 158 F 10 70 C Stora...

Page 11: ...ontact GF Health Products Inc at the address below DO NOT return products to our factory without prior authorization LIMITATIONS AND EXCLUSIONS The foregoing warranty excludes consumables such as filt...