DATA-CUBE³
User Manual
Doc.:
DATA-CUBE-User-Manual
Issue:
1.3
Date:
15-May-2020
© DiGOS Potsdam GmbH
Page 15 / 24
9.
Operation
9.1.
Configuration
This is an example of a default configuration file. It is to be adapted for the experiment and uploaded to the
DATA-CUBE³ via USB before deployment in the field. Please refer to the configuration file manual for a detailed
description of the configuration file.
**********************************************************************
* *
* Configuration File for the 1 & 3 Channel DATA-CUBE *
* *
* Software V0.1 *
**********************************************************************
* Syntax: *
* -
Lines beginning with a star '*' are comments (and are ignored)
*
* - Empty lines are not allowed *
* - Parameter keywords are of six capitalized characters, *
* directly followed by an equal sign *
* - The parameter value starts at column #8 *
* - Line length is limited to 70 characters *
**********************************************************************
****** !!! Parameters for 3 channel Cubes only !!! ******
**********************************************************************
*
** Active Channels 1 = ch1 // 2 = ch1 & ch2 // 3 = ch1 to ch3
CH_NUM=3
*
** Amplifier Gain 1, 2, 4, 8, 16, 32 or 64
P_AMPL=16
*
** AD Converter Mode 0 = Low power // 1 = High resolution
C_MODE=0
*
** Amplifier Chopping 0 = off // 1 = on
A_CHOP=1
*
**********************************************************************
***** General parameters for 1 and 3 channel Cubes *******
**********************************************************************
*
** Project Name Project name of max. 20 characters.
ONLY ASCII!!
E_NAME=N/A
*
** Sample Rate 50, 100, 200, 400, (800@1ch) sps
S_RATE=100
*
** Digital High Pass 0 = off // 1 = on
A_FILT=0
*
** FIR Filter 0 = Linear phase // 1 = Minimum phase
A_PHAS=0
*
** Time Base Correction 0 = off // 1 = PLL // 2 = DIFF
PLL_XO=0
*
** Geographic position lat, lon, alt; to speed up satellite search
*GPS_PO=>SIP+52+013+0000
*
** GPS Mode 0 = cycled // 1 = continuously
GPS_ON=0
*
** GPS & Flush Interval 3 to 60 minutes