3. Product description
3.1 Required accessory
The
EASY
LOG
80NS-…
is configurated, started and read-out via the
EASYBus
-interface.
Therefore the following accessories are required:
Level converter
- RS232 <>
EASYBus
(i.e. EBW 1, EBW 64, EBW 240)
- USB <>
EASYBus
(i.e. EBW 3)
Connecting cable: level converter to
EASY
LOG
GSOFT 40K
(version >7.14):
Windows software to start the logger and read out the logger-data
EASYBus-Configurator
(version >2.0):
for the configuration of the extended settings
3.2
Notes on the logger’s delivery status
The delivered logger is in a “sleep mode”, that means nothing is displayed and the
power consumption is at its minimum. When there is a communication with the software
for the first time, the EASY
LOG
leaves the sleep mode and ›Stop‹ is displayed. Then the
device is ready for operation.
Note:
That sleep mode (nothing is displayed) cannot be restored again.
“Stop“ or “Halt“ are similar to the sleep mode. Both of that operation
modes ensure minimum consumption.
Note:
Logger start over keypad:
before logger start over keypad, logger time must be adjust with
GSOFT 40K.
3.3 Programming of the display range
The display range, position of decimal point and the display unit can be set by the
EASYBus-Configurator
software.
Furthermore, it can be used to read out sensor information (unit type, serial number,
address, etc.) and to program the alarm points of the logger.
3.4 EASYBus connector
Assignment of panel jack
(view on pins)
EASYBus
connection
not connected