There is also a service mode for the trim potentiometer offering the following
possibilities:
a) A check of whether potentiometers are positioned at the calibrated factory settings
b) The pre-adjustment of a new output characteristic curve with connection of a
current measuring device. (A temperature calibrator is not necessary)
c) Specification of a constant value at the current output, e.g. in order to test the
reaction of connection devices.
The trim potentiometers can be used in both normal operation and service
mode.
4.2
Service mode
A corresponding service mode can be selected for each of the two trim
potentiometers via the DIP switches.
4.21 Check of factory settings
- If the selected potentiometer is positioned at the calibrated factory setting,
the operation LED illuminates
red (continuously)
.
- If the potentiometer is above the calibrated factory setting, the operation LED
blinks
red
.
- If the potentiometer is below the calibrated factory setting, the operation LED blinks
green
.
(Since the end value can only be reduced, the green blinking is omitted)