13
Display
Description
(entered values are factory settings)
Security area
0000
Enter password
for safety area
! 0010 !
Change password
Please note
Input
Type K
>
Type J
Thermocouple 2x type J
Display
> with
without
decimal place
0.0
°
C
Sensor correction
Entry of the password for the security level (0010 in condition as delivered)
After an incorrect password has been entered three times, the device
enters a safety timeout for 60 minutes. With the setting 0000 and active
password protection, a jump to the working level takes place.
Change the password
Change the value within the range 0000 ... 9999
Attention:
Write down the password. Resetting to 0010 can only take place in the factory
Input
Type J
Double thermocouple Fe-CuNi
Type K
Double thermocouple NiCr-Ni
Type N
Double thermocouple NiCrSi-NiSi
Type S
Double thermocouple Pt10Rh-Pt
Decimal place (only for Pt100 input and double thermocouple type J)
without
no decimal place
with
one decimal place
Measurement correction
Change of the value in the range -10°C (-10.0°C) ... 10°C (10.0°C)