HD2303
- 22 -
V1.6
INSTRUMENT
SIGNALS
AND
FAULTS
The following table lists all error indications and information displayed by the instrument and
supplied to the user in different operating situations:
Display indications
Explanation
- - -
This appears in the display central line when a temperature only probe is connected.
In the lower line the temperature is shown correctly.
>>>_PRBE_TYPE
type of probe connected
BATT TOO LOW
CHNG NOW
Indication of insufficient battery charge appearing on turning on. The instrument
issues a long beep and turns off. Replace the batteries.
CAL
LOST
Program error: it appears after turning on for a few seconds. Contact the
instrument's supplier.
ERR
This appears if the probe has already been detected by the instrument, but is
disconnected. At the same time an intermittent beep is issued.
FUNC CLRD
max, min and average values cleared
NO_PRBE_ SER_NUM
the connected probe's serial number is absent
OVER
Measurement overflow: indicates that the probe is measuring a value exceeding the
measuring range.
PLS_EXIT >>> FUNC
RES_FOR_FACT ONLY
please exit using ESC >>> function reserved to factory calibration
PRBE_SER #### ####
serial number #### #### of the connected probe
PROB
ERR
A probe with SICRAM module has been inserted when not admissible for that
specific instrument.
PROB COMM LOST
This appears if the probe has already been detected by the instrument, but is
disconnected. At the same time an intermittent beep is issued.
SECT inch2
section in inch
2
SECT m2
section in m
2
SYS
ERR
#
Instrument management program error. Contact the instrument's supplier and
communicate the numeric code # reported by the display.
INSTRUMENT
STORAGE
Instrument storage conditions:
•
Temperature: -25…+65°C.
•
Humidity: less than 90%RH without condensation.
•
Do not store the instrument in places where:
•
humidity is high;
•
the instrument may be exposed to direct sunlight;
•
the instrument may be exposed to a source of high temperature;
•
the instrument may be exposed to strong vibrations;
•
the instrument may be exposed to steam, salt or any corrosive gas.
The instrument case is made of ABS plastic: do not use any incompatible solvent for cleaning.