The following table lists all error indications and information displayed by the
instrument and supplied to the user in different operating situations:
Display indications
Explanation
>>> CAL_MODE
Calibration mode
RTD_PRBE_TYPE
Type of probe connected
1ST_MEAS UP DOWN
Correct the first point using the arrows
/
2ND_MEAS UP DOWN
Correct the second point using the arrows
/
BATT TOO LOW
CHNG NOW
Indication of insufficient battery charge appearing on turning on.
The instrument issues a long beep and turns off. Replace the
batteries.
CAL
LOST
Program error: it appears after turning on for a few seconds.
Contact the instrument's supplier.
CAL FACT
Factory calibration
CAL USER
User calibration
ERR
This appears if the probe has already been detected by the
instrument, but is disconnected. At the same time an
intermittent beep is issued.
FUNC CLRD
Max, min and average values cleared
NEW PROBE DET
This message appears when a probe is inserted into a functioning
instrument. Turn the instrument off and then back on again.
NO_PRBE_ SER_NUM
The connected probe's serial number is absent
OVER or UNDR
Measurement overflow: indicates that the probe is measuring
a value exceeding the measuring range.
PLS_EXIT >>> FUNC
RES_FOR_FACT ONLY
Please exit using ESC >>> function reserved to factory
calibration
PRBE_SER #### #### Serial number #### #### of the connected probe
PROB
ERR
A probe with SICRAM module has been inserted when not
admissible for that specific instrument.
PROB COMM LOST
This appears if the probe has already been detected by the
instrument, but is disconnected. At the same time an
intermittent beep is issued.
SEL MEAS 1/2
Select measurement 1 or 2
SYS
ERR
#
Instrument management program error. Contact the
instrument's supplier and communicate the numeric code #
reported by the display.
INSTRUMENT
STORAGE
Instrument storage conditions:
•
Temperature: -25…+65°C.
•
Humidity: less than 90%RH without condensation.
•
Do not store the instrument in places where:
•
humidity is high;
•
the instrument may be exposed to direct sunlight;
•
the instrument may be exposed to a source of high temperature;
•
the instrument may be exposed to strong vibrations;
•
the instrument may be exposed to steam, salt or any corrosive gas.
The instrument case is made of ABS plastic: do not use any incompatible solvent for
cleaning.