HD2307
- 9 -
V1.7
THE
PROGRAMMING
MENU
To access to the menu press simultaneously the following keys:
The items to be set are listed in this order:
1.
Probe type: the "RTD_PRBE_TYPE" message is displayed in the comment
line. The main line in the center of the display shows the type of probe
connected to the instrument. The following probes can be connected to the
input:
•
temperature probes Pt100 complete with SICRAM module
•
direct 4-wire Pt100 probes complete with TP47 module
•
direct 3-wire Pt100 probes complete with TP47 module
•
2 or 4-wire Pt1000 probes complete with TP47 module
Upon turning on the instrument automatically detects the probes fitted with
SICRAM module: the Probe Type menu item is configured as AUTO and cannot
be modified by the user.
Upon turning on the direct 4 wire Pt100, direct 3 wire Pt100, and 2 wire Pt1000
temperature probes display the "NO_PRBE_SER_NUM" ( no serial number
of the connected probe) message; in this case the probe type must be
entered manually. Select Probe type using the MENU key and then select the
type of probe with the arrow keys; confirm using ENTER.
•
use the
and
arrows (respectively located above the HOLD and REL keys)
to modify the type of probe;
•
press DATA/ENTER to confirm the modification and go onto the next item;
•
press CLR/ESC to cancel the modification;
•
to exit the menu, press the UNIT/MENU key again.
2.
Starting the User calibration procedure: the ">>>_CAL_MODE" message
is displayed in the comment line, and "FACt" is displayed in the main line.
The instrument is provided with the factory ("FACt") calibration. It is also
possible to perform a "USER calibration" ("USEr") of insprobe. The
calibration information is saved in the instrument memory and not in the probe.
The same correction is applied to any probe connected to the input: therefore,
the "USER calibration" should only be used with the probe used for calibration
and not with other probes.
•
use the
and
arrows (respectively located above the HOLD and REL
keys) and select USEr, to access the "USER calibration" procedure;
•
press DATA/ENTER to confirm the modification;
•
the "SEL_MEAS_1/2" message is displayed in the comment line;
•
use the
and
arrows (respectively located above the HOLD and REL
keys) to select "0", "1" or "2" in the main line;
•
press DATA/ENTER to confirm the modification;
•
press CLR/ESC to cancel the modification;
•
to exit the menu, press the UNIT/MENU key again.