WARNINGS
1.
Do not expose the probes to gases or liquids that could corrode the material of the
probe. Clean the probe carefully after each measurement.
2.
Do not bend the probe connectors or force them upward or downward.
3.
Comply with the correct polarity of the probes.
4.
Do not bend or force the contacts when inserting the probe connector into the
instrument.
5.
Do not bend, deform or drop the probes, as this could cause irreparable damage.
6.
Always select the most suitable probe for your application.
7.
In general, do not use temperature probes in presence of corrosive gases or
liquids; the probe external containers are made of AISI 316 or INCONEL stainless
steel, while the contact probe containers are made from AISI 316 or INCONEL
stainless steel plus silver. Avoid contact between the probe surface and any sticky
surface or product that could corrode or damage it.
8.
To obtain reliable measurements, temperature variations that are too rapid must
be avoided.
9.
Temperature probes for surface measurements (contact probes) must be held
perpendicular against the surface. Apply oil or heat-conductive paste between the
surface and the probe in order to improve contact and reduce reading time.
Whatever you do, do not use water or solvent for this purpose. A contact
measurement is always very hard to perform. It has high levels of uncertainty and
depends on the ability of the operator.
10.Temperature measurements on non-metal surfaces usually require a great deal of
time due to the low heat conductivity of non-metal materials.
11.The sensor is not insulated from its external casing; be very careful not to come
into contact with live parts (above 48V). This could be extremely
dangerous for the instrument as well as for the operator, who could be
electrocuted.
12.Avoid taking measurements in presence of high frequency sources, microwave
ovens or large magnetic fields; results may not be very reliable.
13.The instrument is water resistant, but should not be immersed in water. Should
the instrument fall into the water, check for any water infiltration from the
connectors' side. Gently handle the instrument in such a way as to prevent any
water infiltration from the connectors' side.