HD2328
- 9 -
V1.7
THE
PROGRAMMING
MENU
To access to the programming menu press simultaneously the
DATA+(A-B)
keys.
The items to be set are listed in this order:
1.
Selection of the unit of measurement: the "SEL_MEAS_UNIT" message is
displayed in the comment line. The main line in the center of the display shows
the selected unit of measurement: Celsius (°C) or Fahrenheit (°F) degrees.:
•
use the
and
arrows (respectively located above the HOLD and
REL keys) to modify the type of probe;
•
press DATA/ENTER to confirm the modification and go onto the next
item;
•
press CLR/ESC to cancel the modification;
•
to exit the menu, press the A-B/MENU key again.
2.
Selection of the type of Thermocouple: the "SEL" message is displayed in
the main line, while the type of probe is displayed in the comment line; in the
secondary line is shown the "tc" message.
The types of thermocouple that can be selected are: K, J, T, or E.
•
use the
and
arrows (respectively located above the HOLD and
REL keys) to modify the type of thermocouple;
•
press DATA/ENTER to confirm the modification and go onto the next
item;
•
press CLR/ESC to cancel the modification;
•
to exit the menu, press the A-B/MENU key again.
3.
Starting the User calibration procedure: the ">>>_CAL_MODE" message
is displayed in the comment line, and "FACt" is displayed in the main line.
The instrument is provided with the factory ("FACt") calibration. It is also
possible to perform a "USER calibration" ("USEr") of insprobe. The
calibration information is saved in the instrument’s memory. The same
correction is applied to any probe connected to the input: therefore, the "USER
calibration" should only be used with the probe used for calibration and not with
other probes.
•
use the
and
arrows (respectively located above the HOLD and REL
keys) and select USEr, to access the "USER calibration" procedure;
•
press DATA/ENTER to confirm the modification;
•
the "SEL_CHAN" message is displayed in the comment line;
•
use the
and
arrows (respectively located above the HOLD and REL
keys) to select the input "A" or "B" in the main line;
•
the "SEL_MEAS_1/2" message is displayed in the comment line;
•
use the
and
arrows (respectively located above the HOLD and REL
keys) to select "0", "1" or "2" in the main line;
•
press DATA/ENTER to confirm the modification;
•
press CLR/ESC to cancel the modification;
•
to exit the menu, press the UNIT/MENU key again.
Summary of Contents for Delta OHM HD2328.0
Page 4: ...HD2328 4 V1 7 Thermocouple Thermometer HD2328 0 ...
Page 19: ......