HD2010UC
- 11 -
V4.2
•
HD2010PNE2W
: heated preamplifier with connector for ½” UC52 microphone and driver for
extension cables. This preamplifier, supplied with CTC calibration device, can be connected to
sound level meters using the 5mt supplied extension cable (10m as option).
•
HD2010PNE4
: preamplifier for MC24E ¼” microphone. Equipped with CTC calibration device
for electric calibration and driver for cable up to 100m. Requires the HDP079A02 microphone
adapter.
•
HD2010PNE4H
: preamplifier for MC24EH ¼” microphone. Equipped with CTC calibration
device for electric calibration and driver for cable up to 100m. Requires the HDP079A02 micro-
phone adapter.
T
HE
I
NSTRUMENT
The signal of the preamplifier comes to the instrument receiver and its output is sent to the
LINE connector and to the A/D converter input. The instrument can be set to use the LINE channel
in place of the signal coming from the preamplifier.
The A/D converts the analogue signal into the numeric format at 20 bits. The measurement dynam-
ics, exceeding 140 dB, is divided into 5 ranges using a variable gain amplifier with steps of 10 dB,
between 0 dB and 20 dB, positioned at the input.
The signal of the preamplifier comes to the instrument receiver and its output is sent to the
LINE connector and to the A/D converter input. The instrument can be set to use the LINE channel
in place of the signal coming from the preamplifier.
The A/D converts the analogue signal into the numeric format at 20 bits. The measurement dynam-
ics, exceeding 140 dB, is divided into 5 ranges using a variable gain amplifier with steps of 10 dB,
between 0 dB and 20 dB, positioned at the input.
The levels with (A, C and Z) wideband frequency weighting are calculated in parallel in the
DSP). Peak (C and Z) levels are also calculated. The levels calculated by the DSP are transmitted to
the microprocessor for further processing, ready to be displayed, stored and printed.
The microprocessor controls all the instrument processes: management of the electrical calibrator,
Flash memory, display, keyboard and multi-standard serial interface (RS232 and USB).
The microprocessor also supplies the electrical signal corresponding to the A weighted instant level
with FAST time constant, sent to DC output.