HD2010UC
- 16 -
V4.2
before pressing PAUSE, for example, because they were caused by unexpected events and not char-
acterizing the sound being examined.
During measurement, press PAUSE/CONTINUE: integrated measurements update will be in-
terrupted. At this point, press the LEFT arrow to delete the last recorded data.
The integration time value will be temporarily replaced by the word ERASE followed by the
time interval (in seconds) to be deleted. Use the LEFT and RIGHT keys to increase or decrease the
erase interval. Displayed integrated parameters change accordingly, allowing choosing the erase
time depending on the effective need. When pressing PAUSE/CONTINUE again, measurement will
start again and the integrated parameters will have been removed from the selected interval.
The erase maximum time, divided into 5 steps, is set from menu: MENU >> Instrument >>
Measurement >> Max Back-Erase. Settable values are: 5, 10, 30 or 60 seconds, with 1s, 2s, 6s or
12s steps, respectively.