HD2010UC
- 26 -
V4.2
S
INGLE
I
NTEGRATION
When the integration mode is
single
, the sound level meter resets the integrated parameters
(i.e. Leq), starts measuring the instant sound levels (i.e. SPL) and calculates the integrated levels
(i.e. Leq, maximum and minimum levels or statistical levels), regularly until acquisition stops.
This mode provides, at the end of the measurement session, the integrated levels over the
whole acquisition period. The parameter “MENU >> Instrument >> Measurement >> Integration
Interval” allows to suspend the display update, when set time is over. Now, to store displayed data
press REC and hold it down
for at least 2 seconds
, then select the manual storage option. Press
PRINT if you want to send displayed data to the serial output.
While display update is paused (HOLD mode), the sound level meter continues measuring
and calculating the sound levels; press HOLD to let display update start again. If you do not wish to
continue beyond the set integration time, press STOP to interrupt the acquisition.
When the integrating mode is
SINGLE
and the sound level continuous recording is active,
the acquisition is automatically ended once the set integration time is elapsed.
The PAUSE/CONTINUE key can be used to suspend the calculation of integrated levels tem-
porarily, while instant levels are still being measured. During a pause, and as far as the integrated
levels displayed on the SLM page are concerned, you can delete the last integration seconds through
the “Back-Erase Function (data exclusion) ” see page. 15. The monitor function is not affected by
the acquisition pauses. The continuous recording function suspends data storage during the acquisi-
tion pauses and automatically stores a marker indicating the pause duration and the possible use of
the erase function.
The sound level meter has an additional timer for acquisition over time intervals (MENU >> Gen-
eral >> Measures >> Report). With this function it is possible to split the measurement time over
programmable duration intervals form 1 second up to 1 hour; for each interval it is then possible to
calculate a set of 5 integrated levels choosing from Leq, max, min levels, SEL and statistical levels.
These data are not directly available on the display but can be stored activating continuous record-
ing mode. The report levels can be evaluated directly on the sound level meter display using the
“Navigator” program, selecting the Report mode for the replay.
Interval acquisition can be activated only in “SINGLE” integration mode.