HD2010UC
- 40 -
V4.2
If the
Auto-Store
function is activated, SLM screens will be automatically stored when the set in-
tegration time is over. When this function is activated with the multiple integration mode, re-
cording is performed automatically according to the set integration time. Integration time is pro-
grammable from SLM display or using the relative menu item (Menu >> Instrument >> Meas-
urement >> Integration interval). The enabling of the recording mode is indicated by a flashing
REC over the status indicator. Recording is started by pressing RUN. To disable the Auto-Store
function press the REC key briefly.
R
EPORT
In this menu the recording of the reports is defined, (it requires the “Advanced Data Logger”
option). Each item can be enabled separately, as with the measurement recording. In order to
avoid occupying memory space uselessly, we recommend to enable only the needed items and
set the others to OFF. The integration interval (recording time) of the reports is programmable
using the relative menu item (Menu >> Instrument >> Measurement >> Report Time).
•
Par.1 – Par.5
: define five integrated parameters, with relevant frequency weightings.
•
Parameters
: enables recording of the 5 parameters Par.1 – Par.5, defined previously.
•
Statistics
: activates the recording of the statistics.
E
VENT
In this menu (it requires the “Advanced Data Logger” option) the recording of the event re-
ports is defined, as described in the paragraph THE RECORD FUNCTION”.
Each item can be enabled separately, as with the measurement recording. In order to avoid occu-
pying memory space uselessly, we recommend to enable only the needed items and set the others
to OFF.
•
Par. 1 Par.
5: define five integrated event parameters, with relevant frequency weightings.
•
Parameters
: enables recording of the 5 parameters Par.1 – Par.5, defined previously.
•
Statistics
: activates the recording of the event statistics
C
ALIBRATION
•
Calibration Level
: Sound level of the reference sound source used for the sound level meter
calibration. Allowed values vary from 90.0dB to 130.0dB with a 0.1dB resolution.
•
Microphone Response
: allows selecting the type of frequency response of the microphone
according to the sound field. The standard setting for the standard microphone (UC52) is
“Free Field” (FF), since it has a frequency response optimized for the “free field”. Set the pa-
rameter to “Random Incidence” (RI) to activate the correction for random incidence sound.
This setting is necessary to carry out measurements according to ANSI standards. The random
incidence correction is not available on the HD2010UC Class 2.
When HD WME microphone rain protection is used, for aircraft noise (0° incidence) meas-
urements the FF setting must be used while RI setting must be used for community noise (90°
incidence) measurements (e.g. road traffic noise).
•
Wind Shield Correction
: Allows correcting the frequency response of the sound level meter
when the HD SAV windshield provided with the sound level meter, or the HD WME outdoor
shield, is used. When this parameter is set to SAV or WME, the sound level meter frequency
response is corrected for the windshield or the outdoor shield presence, respectively. The ran-
dom incidence correction is not available on the HD2010UC Class 2. For additional details on
applied microphone correction, please see UC2 microphone manual.