HD2010UC
- 60 -
V4.2
SET
G
ROUP
(SETUP)
The following table shows the list of the commands of the SET group (SETUP).
Command
Format
Description
INSTR_MODEL
A10
Instrument model - UNMODIFIABLE
INSTR_NUMBER
A10
Instrument serial number - UNMODIFIABLE
INSTR_VERSION
A10
Instrument version - UNMODIFIABLE
MIC_MODEL
A10
Microphone model – UNMODIFIABLE
MIC_NUMBER
A10
Microphone serial number – UNMODIFIABLE
MIC_TYPE
A10
Type of microphone – UNMODIFIABLE
CLASS
A10
Class of tolerance – UNMODIFIABLE
MEM_SIZE
A10
Memory size – UNMODIFIABLE
OPTIONS
A10
Firmware options – UNMODIFIABLE
TIME
A4
Time (hh:mm)
DATE
A5
Data (aaaa/mm/gg)
DISP_CONTRAST
A8
Display contrast (3÷9, default: 5)
AUTO_POWEROFF
A8
Instrument auto-power-off (ON/OFF, default: ON)
BAUD_RATE
A8
RS232 baud rate
DEVICE
A8
Serial device
INPUT_GAIN
A8
Input gain
INTEGRATION_TIME
A3
Integration time in s, m (1÷59) od h (1÷99)
REPORT_TIME
A8
Report Time
ERASE_TIME
A8
Erase interval
EXCHANGE_RATE
A8
Exchange rate in dB (3÷5)
DOSE_THRESHOLD
A8
Dose threshold in dB (0÷140)
CRITERION_LEVEL
A8
Criterion level in dB (60÷140)
VERBOSE
A8
Acknowledge (ON/OFF, default:ON). Always ON upon power
on.
OVERLOAD_LEVEL
A8
Overload level in dB (20÷200)
INT_MODE
A8
Integration Mode
1_PERC_LEVEL
A8
Percentile level 1 in % (1 ÷ 99, default: 1)
2_PERC_LEVEL
A8
Percentile level 2 in % (1 ÷ 99, default: 10)
3_PERC_LEVEL
A8
Percentile level 3 in % (1 ÷ 99, default: 50)
4_PERC_LEVEL
A8
Percentile level 4 in % (1 ÷ 99, default: 90)
1_SLM_PARAMETER
A6
SLM parameter 1 (see parameter list)
2_SLM_PARAMETER
A6
SLM parameter 2 (see parameter list)
3_SLM_PARAMETER
A6
SLM parameter 3 (see parameter list)
STAT_PARAMETER
A6
Parameter for statistical analysis
EVN_TRIGGER
A8
Event trigger source
EVN_ON_LEVEL
A8
Trigger activation level dB (10 ÷ 140, default: 90)
EVN_OFF_LEVEL
A8
Trigger deactivation level dB (10 ÷ 140, default: 60)
EVN_POLARITY
A8
Trigger level polarity (POS/NEG)
EVN_ON_TIME
A8
Trigger activation delay in seconds from 0 to 10
EVN_OFF_TIME
A8
Trigger deactivation delay in seconds from 0 to 255
EVN_PRINT
A8
Enabling event trigger printing
AUTO_STORE
A8
Enabling Auto-Store function (ON/OFF, default: OFF)
SLM+PROF_DLOGGE
R
A8
Continuous recording of SLM and PROFILE parameters
(ON/OFF, default: ON)
1_REP_PARAMETER
A6
REPORT parameter 1 (see parameter list)
2_REP_PARAMETER
A6
REPORT parameter 2 (see parameter list)