HD2010UC
- 75 -
V4.2
M
ETROLOGICAL
C
HARACTERISTICS
Frequency Weighting
A, C, Z for RMS measurements
C, Z for peak level measurements
Z weighting is flat along the whole sound spectrum with the following features:
Attenuation
[dB]
Frequency Range
[Hz]
< 0.1
100 ÷ 20 k
< 1
31.5 ÷ 21 k
< 5
16 ÷ 22.5 k
Frequency Response
Res
po
ns
e/dB
Frequency / Hz
The selection of the UC52 microphones, on the basis of frequency response specifications,
and the spectral corrections calculated by the sound level meter DSP allows obtaining a Class 1
response according to IEC 61672. The HD2010UC is also available in the Class 2 version.
The frequency response of the microphone depends on the presence of devices like wind-
shield HD SAV or all-weather protection unit HD WME. In order to make measurements with
the maximum possible precision in different situations, the HD2010UC sound level meter can
automatically apply spectral corrections needed to measure in random field conditions, or with
the windshield or with the all-weather outdoor protection. Spectral correction for diffuse field
measurement is enabled setting parameter Menu >> Calibration >> Mic.Resp.Correction >> RI
while correction for HD SAV mounted, is enabled setting parameter Menu >> Calibration >>
Shield >> SAV.
For more information on the applied corrections please refer to the UC52 microphone manual.
HD WME
outdoor protection
The windshield and rain protection units mounted on HD WME outdoor protection, modify
microphone frequency response: in order to apply the correction to sound level meter overall re-
1. Tol
2. Sound level meter response
3. Microphone response
4. Tolerance –