HD2010UC
- 89 -
V4.2
The sound level meter calculates the following parameters:
DOSE (A)
, the percentage of the daily
effective dose, and
DOSE,d (A),
the estimated daily DOSE according to the programmed parame-
ters.
The DOSE calculation is characterized by three parameters:
1.
DOSE Criterion
is the SPL constant value, which continuous exposure for 8 hours determines a
100% DOSE.
2.
DOSE Threshold
that represents the SPL level under which the DOSE is not increased.
3.
Exchange rate
is the variation of the SPL value that determines a double or half duration of ex-
posure with the same DOSE Criterion. Provided values are 3, 4 or 5dB.
The three configuration parameters are contained in the submenu Measurement (MENU >> Instru-
ment >> Measurement): once they have been set, select the submenu Sound Level Meter (MENU
>> Sound Level Meter) and choose, according to the type of measurements to be made, between the
parameters DOSE (A) or DOSE,d (A).
The integration time can be entered directly in the SLM measuring window. Now the instrument is
ready to carry out the measurement: press START. After Tint time has elapsed, the instrument turns
into the HOLD status and displays the DOSE calculated over the set time.
S
TATISTICAL
A
NALYSIS
Up to 3 percentile levels are selectable in the SLM display mode (MENU >> Instrument >>
Measurement >> Percentile Lev. 1-4), programmable between L
1
and L
99.
The statistical analyzer
samples the sound pressure A weighted level with FAST time constant, 8 times a second; the levels
are accumulated in 0.5dB classes. The percentile levels are calculated interpolating the cumulative
distribution.
The “Advanced Data Logger” option can be used to choose which parameter the statistical analysis
is performed on: equivalent level, sound pressure level with FAST time constant, and peak level.
The complete statistical analysis is available with the distribution of probabilities graph and the L
1
to L
99
percentile level graph.
D
ATA
P
RINTING
The displayed values can be printed at any time in all display and acquisition modes.
Besides the
Monitor
function can be activated via the serial line when you press PRINT and hold it
down for 2 seconds. This function allows to continuously send displayed data in real time to the se-
rial interface. The transferred data are those of the active display mode at the time PRINT was
pressed. Data are continuously transferred until PRINT is pressed again, or until the acquisition is
stopped. The Monitor function can be activated even in the STOP acquisition mode; it will start as
soon as the instrument turns into RUN mode. The Monitor works independently from any recording
of data on memory.
Using the Monitor function it is possible, with the help of a PC, to make measurements limited
only by the PC storage capacity.