HD2010UC
- 90 -
V4.2
TROUBLESHOOTING
The HD2010UC sound level meter is provided with a diagnostic program (DIAGNOSTIC
CHECK) that automatically checks the instrument main parameters. This program can be run at any
time to check the instrument operating conditions (see description on page 54).
One of the parameters being analyzed is the sensitivity of the amplification channel that includes,
through a charge partition circuit (CTC), the microphone capacity. Measurement is made at 1kHz.
DIAGNOSTIC CHECK
1.
The DIAGNOSTIC CHECK program fails
Replace batteries and try again after waiting for the end of the stabilization time and, if the
problem remains, contact service.
CALIBRATION
1.
The ELECTRIC CALIBRATION program fails
Ensure that the instrument is not subject to high noise and/or vibrations.
Try again after having waited for the end of the stabilization time and, if the problem re-
mains, run the ACOUSTIC CALIBRATION program.
2.
The ACOUSTIC CALIBRATION program fails
Ensure that the instrument is not subject to high noise and/or vibrations and that acoustic
calibrator and sound level meter are steadily aligned and that the microphone is properly
plugged in the calibrator cavity. Check that the seal rubber ring is present and undamaged.
Try again after having waited for the end of the stabilization time and, if the problem re-
mains, load the factory calibration following these steps:
•
Ensure that acquisition is on STOP.
•
Remove one of the batteries while the instrument is switched on: this operation ensures
that all the instrument internal circuits are discharged.
•
Press and
hold ENTER down
, and then insert the battery you had removed. The instru-
ment will switch on and will display a warning relevant to the load of factory calibration.
Release ENTER and press the key on the right near the CONTINUE key.
•
After having waited for the stabilization time, run the program ACOUSTIC
CALIBRATION.
Should the program fail, contact the service department.
R
ESTORING
F
ACTORY
S
ETUP
The default configuration of the instrument parameters (factory setup) can be recalled at any
time by means of a combination of keys.
This operation does not clear the content of data
memory
.
While the instrument is off, press and hold down ENTER and power on the sound level meter. All
the menu items are simultaneously brought back to the default value.
R
ESTORING
F
ACTORY
C
ALIBRATION
The Factory Calibration can be recalled at any time by means of a combination of keys.
This
operation does not clear the content of data memory
.
While the instrument is off, remove one of the batteries and wait at least 5 minutes for the sound
level meter internal circuits to discharge.