HD2010UC
- 95 -
V4.2
PAUSE/CONTINUE key
The PAUSE key interrupts the calculation of integrated measurements (Leq, SEL, maximum or
minimum levels, spectra, etc.), as well as recording. The instantaneous levels are still measured and
displayed in the SLM screen. Press PAUSE/CONTINUE to start measuring again. The integrated
parameters are cleared if you press RUN/STOP/RESET while in PAUSE (during a measurement
session). The last seconds of integration can be excluded, from the calculation of integrated parame-
ters shown on the SLM screen (for example, to eliminate the effect of an undesired noise), using the
LEFT and RIGHT keys while in PAUSE. The maximum erasing interval is programmable from 5
seconds to 60 seconds via MENU >> Instrument >> Measurement.
If you press the RUN/STOP/RESET key while in pause during the replay of a recording (with the
“Advanced Data Logger” option), the next stored data will be displayed. If you press the
RUN/STOP/RESET key and hold it down, the replay will be performed in fast forward mode.
REC key
If REC is pressed and hold down for at least 2 seconds, the displayed data are stored as a single re-
port. You can also activate the automatic recording of the parameters displayed on the SLM screen,
(see THE RECORD FUNCTION on page 30).
When the “Advanced Data Logger” option is installed, the REC key, combined with
START/STOP/RESET, activates the data logging.
Starting from the STOP condition
, if you press
REC and hold it down, then press START/STOP/RESET, the data logging of measured values is
started. To stop logging, press START/STOP/RESET: registration number, date and time will be
displayed. Press ENTER to confirm.
RUN/STOP/RESET key
If you press RUN, while in stop, all the initial values of the integrated measurements like Leq, SEL,
MAX/MIN levels will be first cleared (RESET) and then measurement starts. Pressing the key
again (STOP), the measurements will be stopped.
If pressed while in pause, all integrated pa-
rameters will be cleared.
During the replay of stored data (with the “Advanced Data Logger” option), if you press this key
while in pause, the next data will be displayed; if you press it and hold it down, the replay will be
executed in fast forward mode.