HD2010UC
- 34 -
V4.2
Delayed acquisition timer
A timer is available to enable data acquisition according to a programmable delay of up to 99
hours. To perform an acquisition with delayed start, the recording parameters need to be set first
and then the
delayed acquisition timer
programmed in the parameter Menu >> Sequencer >> Tim-
er.
After programming the sound level meter, you only need to press the
REC
and
RUN
keys together
(as if starting a logged measurement): confirm with the “YES” key. The instrument switches into
stand-by and turns off (see. Fig. 12).
Fig. 12 - Delayed acquisition timer warning screen
The instrument will turn back about one minute before the set delay, to allow execution of the
warm up time before starting the automatic acquisition. During this waiting minute, the “TIMER”
message blinks, showing the automatic acquisition feature has been activated.
Fig. 13 - Waiting for the timer star
The acquisition will end when the
set integration time (Tint) has elapsed
and the instrument will
automatically turn off after disabling the timer.