E40.0.6x.6C-01 

Operating Manual EASY

LOG

 80BIN 

page 14 of 14 

 

11.  Device options 

Additi

onal information’s for options of the 

EASY

LOG

 80BIN

 

11.1 ALARM 

With the o

ption „ALARM“ the device has an additional switching output (NPN) for  

the alarm function of the device. 
In result of the option there are following extensions: 

Device connection: 

Alarm output 

assignment of jack 

(view on pins) 

GND 

NPN 

 

 

Note to included alarm 
connection cable: 

GND:  back + blue 

NPN:  brown + white 

 

 

EASYBus 

assignment of jack 

(view on pins) 

 

EASYBus 

 

 

 

 

not used

 

Connection examples: 

a.)  General connection diagram for a relay (with self-induction recuperation diode): 

      

 

b.)  Connection information for GR 10, GNR 10 or GNR 232 A: 

GND:

  to be connected to terminal 11 of the 

GNR… 

NPN:

  to be connected to terminal 10 (or. 9) of the GNR... 

Specification: 

Alarm function: 

alarm points selectable within the measuring range 
and alarm delay adjustable from 0 to 500 min. 

Alarm: 

via display (AL.xx), interface communication and 
alarm output (transistor is conducting at alarm fault) 

Alarm output: 

open-collector 

max. switching power: 

28 V  /  50 mA 

Attention:  the alarm output is not electrically isolated 

 

When bus operating of several data loggers with option ALARM a  
electrical connection between the alarm outputs is not allowed!! 

EASY

LOG

 

alarm output 

NPN 

GND 

power supply 

(max. 28V)

 

-Ub 

+Ub 

Summary of Contents for Greisinger EASYLOG 80BIN

Page 1: ...RMANY Tel 49 9402 9383 0 info greisinger de www greisinger de Installation and operating manual State data logger from version 2 4 EASYLOG 80BIN Please read these instructions carefully before use WEE...

Page 2: ...es at the device logger operation 7 5 3 Operating possibilities via software GSOFT 40K 8 5 4 Operating possibilities via software EASYBus Configurator 8 6 Battery life time and recording period 8 7 Op...

Page 3: ...and consequential damages are voided with misuse disregard of this operating manual assignment of inadequately qualified technical personnel and arbitrary modifications of the product 2 2 Safety signs...

Page 4: ...only be guaranteed if the device is not subjected to any other climatic conditions than those stated under Specification 5 When connecting the device to other devices e g PC the interconnection has t...

Page 5: ...he extended settings 3 2 Notes on the logger s delivery status The delivered logger is in a sleep mode that means nothing is displayed and the power consumption is at its minimum When there is a commu...

Page 6: ...ssibilities at the device display selection Function Procedure Display max value Push button shortly The max value is displayed for 4 seconds The max value display function is signalled by the display...

Page 7: ...pressed for more than 2 min the menu is quit without saving Start recording Push button for about 5 seconds The main display shows run The secondary display shows no To start recording select YES with...

Page 8: ...ing possibilities via software EASYBus Configurator The software EASYBus Configurator V2 0 is needed to change the configuration settings With that software the following adjustment are additionally p...

Page 9: ...tatus messages are dis played STOP The EASYLOG recording has been stopped The logger memory is empty Note At this state the power consumption is at its minimum HOLD The EASYLOG recording has been halt...

Page 10: ...r to the manufacturer ERROR 7 The EASYLOG has detected a system error remedy Reset the error via GSOFT 40K If the error message is displayed furthermore please send the logger to the manufacturer for...

Page 11: ...to 5 hours 18000 seconds Memory type cyclic logger recording can be activated by start stop function i e key press boundaries 64 independent recording sets are possible ring logger The logger records...

Page 12: ...t and disposal 9 1 Reshipment All devices returned to the manufacturer have to be free of any residual of measuring media and other hazardous substances Measuring residuals at housing or probe may be...

Page 13: ...o not stop active logger has been activated at the configuration of the logger Remedy Deactivate this function by means of EASYBus configurator Problem The alarm settings cannot be changed Possible re...

Page 14: ...ction diagram for a relay with self induction recuperation diode b Connection information for GR 10 GNR 10 or GNR 232 A GND to be connected to terminal 11 of the GNR NPN to be connected to terminal 10...

Reviews: