E40.0.6x.6C-01 

Operating Manual EASY

LOG

 80BIN 

page 3 of 14 

 

2. Safety 

2.1  Intended Use 

The logger 

EASY

LOG 

80BIN 

is designed for long term state monitoring. 

The combination of low power consumption with high battery capacity ensure a long record-
ing time. 

The last 250.000 measuring values are stores in the memory.  
The LCD-display informs about measured state and about the operating status of the logger. 

Only technically qualified persons are permitted to carry out installation, commissioning, 
operation and decommissioning. The qualified personnel must have carefully read and 
understood the operating manual before beginning any work.  

This device must not be used at potentially explosive areas, as safety or emergency stop 
devices or in any other application where failure of the product could result in personal 
injury or material damage. 
This device must not be used at a patient for diagnostic or other medical purpose. 

The liability and warranty of the manufacturer for damages and consequential damages 
are voided with misuse, disregard of this operating manual, assignment of inadequately 
qualified technical personnel and arbitrary modifications of the product. 

2.2  Safety signs and symbols 

Warnings are labelled in this document with the followings signs: 

 

Caution!

  

This symbol warns of imminent danger, death, serious injuries and signifi-
cant damage to property at non-observance 

 

Attention!

  

This symbol warns of possible dangers or dangerous situations which can 
provoke damage to the device or environment at non-observance.  

 

Note!

  

This symbol point out processes which can indirectly influence operation or 
provoke unforeseen reactions at non-observance. 

2.3  Skilled personnel 

are persons who are familiar with the set-up, installation, commissioning and operation of 
the product and have appropriate qualification for their work. For example: 

 

•  Training or instruction and/or authorisation, power circuits and devices/systems in ac-

cordance with the standards of safety for activation, deactivation, disconnection, earth-
ing and identification. 

 

•  Training or instruction in accordance with the standard of safety technology for care 

and use of suitable safety equipment. 

 

 

Summary of Contents for Greisinger EASYLOG 80BIN

Page 1: ...RMANY Tel 49 9402 9383 0 info greisinger de www greisinger de Installation and operating manual State data logger from version 2 4 EASYLOG 80BIN Please read these instructions carefully before use WEE...

Page 2: ...es at the device logger operation 7 5 3 Operating possibilities via software GSOFT 40K 8 5 4 Operating possibilities via software EASYBus Configurator 8 6 Battery life time and recording period 8 7 Op...

Page 3: ...and consequential damages are voided with misuse disregard of this operating manual assignment of inadequately qualified technical personnel and arbitrary modifications of the product 2 2 Safety signs...

Page 4: ...only be guaranteed if the device is not subjected to any other climatic conditions than those stated under Specification 5 When connecting the device to other devices e g PC the interconnection has t...

Page 5: ...he extended settings 3 2 Notes on the logger s delivery status The delivered logger is in a sleep mode that means nothing is displayed and the power consumption is at its minimum When there is a commu...

Page 6: ...ssibilities at the device display selection Function Procedure Display max value Push button shortly The max value is displayed for 4 seconds The max value display function is signalled by the display...

Page 7: ...pressed for more than 2 min the menu is quit without saving Start recording Push button for about 5 seconds The main display shows run The secondary display shows no To start recording select YES with...

Page 8: ...ing possibilities via software EASYBus Configurator The software EASYBus Configurator V2 0 is needed to change the configuration settings With that software the following adjustment are additionally p...

Page 9: ...tatus messages are dis played STOP The EASYLOG recording has been stopped The logger memory is empty Note At this state the power consumption is at its minimum HOLD The EASYLOG recording has been halt...

Page 10: ...r to the manufacturer ERROR 7 The EASYLOG has detected a system error remedy Reset the error via GSOFT 40K If the error message is displayed furthermore please send the logger to the manufacturer for...

Page 11: ...to 5 hours 18000 seconds Memory type cyclic logger recording can be activated by start stop function i e key press boundaries 64 independent recording sets are possible ring logger The logger records...

Page 12: ...t and disposal 9 1 Reshipment All devices returned to the manufacturer have to be free of any residual of measuring media and other hazardous substances Measuring residuals at housing or probe may be...

Page 13: ...o not stop active logger has been activated at the configuration of the logger Remedy Deactivate this function by means of EASYBus configurator Problem The alarm settings cannot be changed Possible re...

Page 14: ...ction diagram for a relay with self induction recuperation diode b Connection information for GR 10 GNR 10 or GNR 232 A GND to be connected to terminal 11 of the GNR NPN to be connected to terminal 10...