Temperature unit
UN,T
°
[
Display temperature in °C
°
F
Display temperature in °F
Restore factory settings
IN,T
NO
Keep current configuration
YES
Load factory settings (display: “
IN,T DONE
“)
During menu exit, the changes are stored (“
STOR
“) – if necessary the device will be restarted automatically.
8 Adjustment of temperature input
Invoke menu:
9
10
Hold the key “down“during
switching on
the device until “
T.OF
“is shown in the auxiliary display.
The menu can be used like described in chapter 7 Configuration.
Parameter Value
Description
/
T.OF
Zero point adjustment of the temperature measurement
0.00
(
OFF
)
No zero point adjustment for temperature measurement
-5.00
…
5.00
Offset of temperature measurement in °C
(or -
9.00 … 9.00 °F)
T.SL
Slope adjustment of temperature measurement
0.00
(
OFF
)
No slope adjustment for temperature measurement
-5.00
…
5.00
Slope correction of temperature measurement in [%]
The temperature input can be adjusted with offset (
T.OF
) and scale (
T.SL
). A reasonable adjustment presumes
reliable references (e.g. ice water, controlled precision water bath, etc.).
If the inputs are adjusted (i.e. offset and scale are different from factory settings) the device will shortly
display “
T.OF
“/ “
T.SL
“ after turned on.
Default setting for offset
and scale are “
OFF
“=
0.00
, i.e. inputs are not changed.
Zero point correction:
Displayed value = measured value
–“
T.OF
”
Zero point and slope correction:
Displayed value = (measured value -
T.OF
)
• (1 +
T.SL
/ 100)
Displayed value °F = (measured value °F - 32°F -
T.OF
)
• (1 +
T.SL
/ 100)
Example:
Before starting the measurement, the values for “
T.OF
”
and “
T.SL
” are set to “
0.00
(
OFF
)” – “°
[
“ is
selected
for the parameter “
UN,T
”. For reference in a water bath, a clinical thermometer is
used.
Display value in ice water (given value offset = 0.0 °C):
Display val. offset = -0.2 °C
Display value in water bath (given value slope = 37.0 °C):
Display val. slope = 36.6 °C
Calculation:
T.OF
= display value offset - given value offset= -0.2 °C
– 0.0 °C = -0.2 °C
T.SL
= (given value slope / (display val. slope -
T.OF
) - 1) * 100
= (37.0 °C / (36.6 °C - (-0.2)) - 1) * 100
≈ 0.54 (rounded)