background image

H56.0.02.6C-01 

Operating Manual GTH 1150  

Page 6 von 10 

 

 

_____________________________________________________ 

_____________________________________________________________________________ 

 

Adjustment 

The device is calibrated and adjusted at the factory so that any standard probe offered by Greisinger can be 
plugged in without recalibration. If the value deviates excessively over time, we offer factory re-calibration. 

4.1  Manual Adjustment 

 

Manual adjustment changes the factory settings, which can no longer be simply restored! 

If you want to adjust the device to the available probe, the probe must be plugged in first. Then the zero-
point potential must be adjusted. See chapter 

Fehler! Verweisquelle konnte nicht gefunden werden.

 

Connections. 

 

Ensure that the thermometer has adjusted to room temperature for the zero-point 
potentiometer adjustment! 
  

4.1.1  Zero point-Adjustment 

Put ice cubes in a glass and pour cold water until ice cubes are almost covered. Put sensor into glass, wait 
approx. 30 minutes, then stir water with a spoon handle. Wait for stable value to be displayed, and then turn 

zero point potentiometer by means of a screwdriver until display shows 

0.

 

4.1.2  Scale Adjustment 

An additional temperature reference is required for the scale adjustment. Adjust the display with the gradient 
potentiometer (scale) to the desired reference value. 

5  Basics of the measurement 

5.1  Probe accuracy/device accuracy 

DIN EN 60584-2 

Class 

Type 

Deviation 

Validity 

+/- 1.5 °C 

-

40… +375 °C 

5.2  Possible errors 

5.2.1  Immersion depth  

For measurements in liquids the probe should be immersed sufficiently deep (depending on probe 
diameter, at least 20 mm with Ø 3 mm and 10 mm with Ø 1.5 mm) and subsequently stirred.  

When measuring gases the probe should also emerge as deep as possible in the gas to be 
measured (e.g. when measuring in channels/pipes) and the gas should flow around the probe at 
sufficient flow.  

5.2.2  Surface effects and bad heat transfer 

For surface temperature measurements special surface probes are necessary. Surface quality, heat 
transfer and ambient temperature have an influence on the measurement result. 

 

Thermally conductive paste between the probe and surface can also increase 
measurement accuracy in some cases. 

5.2.3  Cooling (evaporation) 

For air temperature measurements, the probe should be dry, otherwise it can be possible that a too 
low temperature could be measured. 

5.2.4  Response time 

Before reading the measured value at the measuring process, it is necessary to wait a sufficient time 
(see chapter 6 Type K-sensor). 

 

Summary of Contents for GREISINGER GTH 1150

Page 1: ...tauf GERMANY Tel 49 9402 9383 0 info greisinger de www greisinger de Operating manual Quick Response Thermometer As of version 1 0 GTH 1150 Please carefully read these instructions before use WEEE Reg Nr DE 93889386 Please consider the safety instructions Please keep for future reference ...

Page 2: ...EMENT 5 3 2 PUSHBUTTONS 5 3 3 CONNECTIONS 5 4 ADJUSTMENT 6 4 1 MANUAL ADJUSTMENT 6 4 1 1 Zero point Adjustment 6 4 1 2 Scale Adjustment 6 5 BASICS OF THE MEASUREMENT 6 5 1 PROBE ACCURACY DEVICE ACCURACY 6 5 2 POSSIBLE ERRORS 6 5 2 1 Immersion depth 6 5 2 2 Surface effects and bad heat transfer 6 5 2 3 Cooling evaporation 6 5 2 4 Response time 6 5 2 5 Limit values 7 6 TYPE K SENSOR 7 6 1 DETAILS AB...

Page 3: ...lug connection suitable for all type K thermocouple probes Refer to chapter 6 Type K sensor to view a selection of the available probes The safety requirements see below have to be observed The device must be used only according to its intended purpose and under suitable conditions Use the device carefully and according to its technical data do not throw it strike it Protect the device from dirt 1...

Page 4: ...to when using the device If there is a risk whatsoever involved in running it the device has to be switched off immediately and to be marked accordingly to avoid re starting Operator safety may be a risk if there is visible damage to the device the device is not working as specified the device has been stored under unsuitable conditions for a longer time In case of doubt please return device to ma...

Page 5: ...or a longer period of time Treat device and probes carefully Use only in accordance with above specification do not throw hit against etc Protect plug and socket from soiling Make sure that the device and the probe plug are always exposed to the same temperature Therefore do not hold the probe plug in your hand too long and do not expose the device to an additional heat source otherwise faulty mea...

Page 6: ... 4 1 2 Scale Adjustment An additional temperature reference is required for the scale adjustment Adjust the display with the gradient potentiometer scale to the desired reference value 5 Basics of the measurement 5 1 Probe accuracy device accuracy DIN EN 60584 2 Class Type Deviation Validity 1 K 1 5 C 40 375 C 5 2 Possible errors 5 2 1 Immersion depth For measurements in liquids the probe should b...

Page 7: ...er the permissible limits of your probe must be observed when measuring high temperatures See chapter 6 Type K sensor 6 Type K sensor 6 1 Details about the sensors Application Temperature range Response time T90 Sensor Insertion sensor 65 550 C 40 s GF 1TK T3 Einstechfühler für weichplastische Medien 65 550 C 12 s GF 1TK E1 5 65 550 C 40 s GF 1TK E3 65 550 C 3 s GES 130 Surface senosr 65 900 C 2 s...

Page 8: ...st or in the worst case exploding Ensure that the device is switched off when replacing the battery The switch must be in Position 0 see chapter Fehler Verweisquelle konnte nicht gefunden werden Fehler Verweisquelle konnte nicht ge funden werden and turn it over so that the rear side is facing upwards 1 Open the cover of the device by carefully pushing it off of the device 2 Pull the battery out o...

Page 9: ...obe connected Check is a suitable probe connected 9 Reshipment and disposal 9 1 Reshipment All devices returned to the manufacturer have to be free of any residual of measuring media and other hazardous substances Measuring residuals at housing or sensor may be a risk for persons or environment Use an adequate transport package for reshipment especially for fully functional devices Please make sur...

Page 10: ...e adjustment time of approx 15 minutes has to be taken into account Atmospheric humidity 0 80 RH not condensing Storage temperature 25 70 C Over 50 C please remove battery Display approx 13 mm high 3 digit LCD Power supply 9V Battery type IEC 6F22 Battery service life 1000 hours Low battery warning BAT displayed automatically in case of low battery Dimensions of case 106 x 67 x 30 mm L x B x D Wei...

Reviews: