Important!!  redox - electrodes are very sensitive. Please read manual of the electrode carefully before usage. 

Error and System Messages [only with option display] 

Display 

Description 

Possible fault cause 

Remedy 

FE  1 

Measuring values 
exceeding measuring 
range 

redox value is out of the 
transducers measuring range 

Replace electrode. 

FE  2

 

Measuring values below 
measuring range 

redox electrode defective 

Replace electrode. 

transducer defective 

Return to manufacturer for repair 

FE  7 

System fault 

Error in transducer 

Disconnect transducer from supply and 
reconnect. If error remains: Return to 
manufacturer for repair . 

8.8.8.8 

Segment test 

The transducer performs a display test for 2 seconds after switch on.  
After that it will change the display of the measuring. 

Input signal is not valid 

redox electrode not connected  Check the connection and cable. 

Permissible input range is 
exceeded 

Check if not a wrong electrode (pH or 
others) is connected. 

Replace electrode. 

 