INSTRUCTION MANUAL CENT series balances 60.50.670 3rd issue july 2005
Page 34 di 40
9.2.2 CALIBR
The CENT balance may be calibrated with an external mass.
9.2.2.1 CAL OFF
Both calibration are disabled
9.2.2.2 CAL ON
Both calibration are enabled
9.2.2.3 CAL AUTO
With this selection the calibration is enabled and a timer activates the internal
calibration every 3 hour. The automatic calibration is delayed if, during last 10
minutes, the balance has been operated or serial comand has been issued.
9.2.3 A-ZERO
If the Auto-Zero function is active and when no weights are on the pan, the automatic zero tracking
correction is active: the balance remains at a net value equal to zero. The function is useful to cancel small
drift of the zero reading.
9.2.3.1 ZERO OFF
The zero tracking function is inactive.
9.2.3.2 ZERO ON
The function is active and small changes in zero reading are corrected.
9.2.4 FILTER
The setting of the filter menu controls the dynamic response of the balance.
Selections are.
9.2.4.1 SLO
slow response applicable to disturbed environments (high accuracy weighing)
9.2.4.2 AVG
average value for both high accuracy and
9.2.4.3 FAST
fast response applicable to quickly changing weights (dosing)
9.2.5 SET DEF
All settings, changed by the setup menu are valid as soon as they are changed.
However if the balance is powered off they are lost and when powered up the balance will use different
parameters.
To maintain a custom configuration that will be active at power-on the function SET DEF is implemented.
9.2.5.1 STORE
current settings are stored as user default
9.2.5.2 END
user default settings are left unchanged