10. Set the AM Sensitivity of the UUT to 90% and repeat Step 7 to Step 8 for each of the test frequen-
cies in Datasheet 6, recording the AM depth for each measurement in the appropriate column.
4.3.3.1.4 Procedure - Amplitude Modulation Bandwidth
1. Set the UUT for the following settings:
Level:
0 dBm
Frequency:
10 MHz
AM Sensitivity:
30%
External AM State:
On
2. Set the Spectrum Analyzer to the following settings:
Reference Level
+5 dBm
Span
20 kHz
Sweeptime
400 msec.
Resolution Bandwidth: < 100 Hz
Video Bandwidth:
> 30 kHz
3. Connect the test equipment and UUT as shown in Figure 4-5,
4. Set the Function Generator for a 1 kHz Sine Wave output at 2.00 V peak-to-peak.
NOTE: Measure the output of the Function Generator using a precision meter or oscilloscope to
ensure that its output is 2.00 Vp-p.
5. Set the Spectrum Analyzer to the test frequency. Center the signal using Peak Search and Marker to
Center Frequency functions.
6. Measure and record the difference, in dB, between the carrier amplitude and one sideband ampli-
tude in the first column of Datasheet 7.
7. Vary the rate of the function generator within the specified rate range for the AM specification.
Identify the lowest sideband amplitude. Record the value in the second column of Datasheet 7.
8. Subtract the reading in column 2 from the reading in column 1 and record the value in column 3.
The difference between the two readings should be 3 dB or less.
9. Repeat steps 5 through 8 for the remaining test frequencies in Datasheet 7.