Front Panel Operating Tasks
2500A Series Operation Manual, 34172 Revision C, March 2008
2. Connect the output of the crystal detector to the Ext ALC In connector on the rear panel using an
appropriate cable.
3. Press the System button to invoke the System menu on the display. If the System Menu 1 is dis-
played, press the System 2 Menu soft key to display the System Menu 2.
4. Select the ALC menu item in the System 2 menu, and select the positive or negative detector setting
using the Step Up/Step Down buttons depending on the type of crystal detector used of external
ALC.
5. Select the ALC menu item in the System 2 menu, and select the positive or negative detector setting
using the Step Up/Step Down buttons depending on the type of crystal detector used of external
ALC.
6. Press CW button to invoke the CW menu on the display. Select the "Frequency" menu item in the
CW menu, and enter the desired CW frequency using the numeric keypad, Step Up/Step Down but-
tons, or the rotary knob.
NOTE:
The Power Offset and Power Slope functions are no longer available.
7. Select the "Level" menu item in the CW menu and enter the desired output level using the keypad,
Step Up/Step Down buttons, or the rotary knob.
8. When entering a new level setting using the numeric keypad, use the dBm or dB Units buttons.
Units are assumed to be in dBV.
9. If the LED indicator that is above the RF ON button is not lit, press the RF ON button to enable the
signal at the RF output connector.
The External Power Meter ALC procedure follows:
1. Connect the sensor of the power meter to the sample port of the power splitter or directional coupler.
2. Connect the Analog Out output of the Power Meter to the EXT ALC connector on the rear panel of
the instrument.
3. Adjust the Analog Out range of the power meter from 0.0005 to 2 volts. The power range is
adjusted according to the user's specific needs.
4. Press the System button to invoke the System menu on the display. If the System Menu 1 is dis-
played, press the System 2/2 soft key to display the System Menu 2.
5. Select the ALC menu item in the System 2 menu, and select the Power Meter setting using Step Up/
Step Down buttons.
6. Press the CW button to invoke the CW menu on the display.
7. Select the "Frequency" menu item in the CW menu, and enter the desired CW frequency using the
numeric keypad, Step Up/Step Down buttons, or the rotary knob.